Model 2520 User’s Manual
Range, Filter, and Math
6-3
Maximum readings
The full scale input for each measurement range is shown in
.
Input levels that exceed the maximum levels cause the “Oflow” message to be displayed.
Setting the measurement range
Press the LASER V
L
, DETECTOR 1 I
PD
or DETECTOR 2 I
PD
key, then use the RANGE
▲
and
▼
keys to select a range.
NOTE
Use the lowest range possible without causing an overflow to ensure best accu-
racy and resolution.
Source ranging
As summarized in
, the laser diode current source has two ranges. To set the
range, press the LASER I
L
and EDIT keys, then use the RANGE
▲
and
▼
keys.
Remote range programming
summarizes the commands necessary to control range. See
, for more details on these commands.
Table 6-3
Laser diode current source ranges
Current
range
Resolution
Maximum
output
500mA
10
µ
A
500mA
5A
100
µ
A
5.0A*
* Pulse mode only. 1A maximum in DC mode.
Table 6-4
Range commands
Commands
Description
[:SENSe[1]]:VOLTage:RANGe <range>
:SENSe2:CURRent:RANGe <range>
:SENSe3:CURRent:RANGe <range>
:SOURce[1]:CURRent:RANGe <range>
Set laser diode voltage measure range (5 or 10).
Set photodiode 1 current measure range (0.01, 0.02, 0.05, or 0.1).
Set photodiode 2 current measure range (0.01, 0.02, 0.05, or 0.1).
Set laser diode current source range (0.5 or 5).
Содержание 2520
Страница 250: ...A Specifications ...
Страница 254: ...B StatusandErrorMessages ...
Страница 262: ...C DataFlow ...
Страница 266: ...D IEEE 488BusOverview ...
Страница 281: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 285: ...F MeasurementConsiderations ...
Страница 308: ...G GPIB488 1Protocol ...
Страница 312: ...H ExamplePrograms ...
Страница 317: ...I ContinuousPulseMode ...