14-8
SCPI Command Reference
Model 2520 User’s Manual
Command
Description
Default
parameter SCPI
:SENSe3
:CURRent[:DC]
:RANGe
[:UPPer] <n>
[:UPPer]?
:POLarity <name>
:POLarity?
:AVERage
:COUNt <n>
:COUNt?
[:STATe] <b>
[:STATe]?
:SWEep
:CABort
[:LEVel]<n>
[:LEVel]?
:STATe<b>
:STATe?
:DATA[:LATest]?
Sense 3 subsystem to control detector #2 current measurement:
Path to configure current:
Configure measurement range:
Select range by specifying the expected current reading;
0 to 0.105 (Ranges 0.01, 0.02, 0.05, 0.1).
Query range.
Select polarity. Name = POSitive or NEGative.
Query polarity.
Path to configure digital filter:*
Specify filter count; 1 to 100.
Query filter count.
Enable or disable digital filter.
Query state of digital filter.
Path to sweep abort commands:
Set detector #2 current level for sweep abort (0 to 0.105).
Query detector #2 sweep abort current threshold.
Enable or disable detector #2 sweep abort function.
Query state of detector #2 sweep abort function.
Query latest reading only when trigger model is in IDLE.
0.1
POS
10
OFF
0.105
OFF
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
Table 14-5 (continued)
SENSe command summary
* :AVERage filter commands are global and affect all three measurements (:SENS1, :SENS2, and :SENS3) simultaneously.
Содержание 2520
Страница 250: ...A Specifications ...
Страница 254: ...B StatusandErrorMessages ...
Страница 262: ...C DataFlow ...
Страница 266: ...D IEEE 488BusOverview ...
Страница 281: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 285: ...F MeasurementConsiderations ...
Страница 308: ...G GPIB488 1Protocol ...
Страница 312: ...H ExamplePrograms ...
Страница 317: ...I ContinuousPulseMode ...