Model 2520 User’s Manual
Source-Measure Concepts
5-5
Figure 5-3
Front panel staircase sweep mode pulse parameters
shows how various front panel programming parameters control various stair-
case sweep DC waveform aspects. These parameters are programmed by pressing the
appropriate key:
•
Pulse amplitudes: determined by the programmed START, STOP, and STEP (or #
POINTS) parameters in the CONFIG SWEEP menu.
•
DELAY + PW: determines the time at each step.
Figure 5-4
Front panel staircase sweep mode DC parameters
PW
Delay
Start
Low
Step
Stop
0A
Returns to 0A
because pulse is
less than Low.
PW
(Readings
Taken
and
Averaged)
Delay
(No
Readings
Taken)
Start
Step
Stop
0A