14-28
SCPI Command Reference
Model 2520 User’s Manual
Select laser diode voltage measurement range
[:UPPer] <n>
[:SENSe[1]]:VOLTage[:DC]:RANGe[:UPPer] <n>|UP|DOWN Select laser diode voltage range
Parameters
<n> =
0 to 10.5
Expected reading in volts
DEFault
10
MINimum
0
MAXimum
10.5
UP
Select next higher measurement range
DOWN
Select next lower measurement range
Query
:RANGe?
Query measurement range
:RANGe? DEFault
Query *RST default range
:RANGe? MINimum
Query lowest range (returns 0)
:RANGe? MAXimum
Query highest range
Description
This command is used to select the measurement range for the laser
diode voltage measurement. The range is selected by specifying the
expected reading. The instrument will then go to the most sensitive
reading that will accommodate that reading. For example, if you expect
a reading of approximately 7V then simply let <n> = 7 in order to select
the 10V range.
You can also use the UP and DOWN parameters to select range. Each
time UP or DOWN is sent, the next higher or lower measurement range
is selected. When on the maximum range, sending UP is a No-Op (no
operation). When on the lowest range, sending DOWN is a NO-Op.
Select detector current measurement range
[:UPPer] <n>
:SENSe2:CURRent[:DC]:RANGe[:UPPer] <n>|UP|DOWN
Select detector #1 current range
:SENSe3:CURRent[:DC]:RANGe[:UPPer] <n>|UP|DOWN
Select detector #2 current range
Parameters
<n> =
0 to 0.105
Expected reading in amps
DEFault
0.1
MINimum
0
MAXimum
0.105
UP
Select next higher measurement range
DOWN
Select next lower measurement range
Query
:RANGe?
Query measurement range
:RANGe? DEFault
Query *RST default range
:RANGe? MINimum
Query lowest range (returns 0)
:RANGe? MAXimum
Query highest range
Содержание 2520
Страница 250: ...A Specifications ...
Страница 254: ...B StatusandErrorMessages ...
Страница 262: ...C DataFlow ...
Страница 266: ...D IEEE 488BusOverview ...
Страница 281: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 285: ...F MeasurementConsiderations ...
Страница 308: ...G GPIB488 1Protocol ...
Страница 312: ...H ExamplePrograms ...
Страница 317: ...I ContinuousPulseMode ...