Model 2520 User’s Manual
SCPI Command Reference
14-33
NOTE
The sourcing mode will default to FIXed whenever the Model 2520 goes to the
local state.
Select source function
FUNCtion[:SHAPe] <name>
:SOURce[1]:FUNCtion[:SHAPe] <name>
Select laser diode source function
Parameters
<name> =
DC
Select DC source function
PULSe
Select pulse source function
Query
:FUNCtion[:SHAPe]?
Query source function
Description
This command is used to select the source function for the laser diode
current source. The two functions are:
DC
— With this source function, the source will output a DC level with
the amplitude determined by the :SOUR1:CURR:AMPL command (see
PULSe
— In this mode, the source will output a current pulse. The high
and low amplitudes are set with the :SOUR1:CURR:AMPL and
:SOUR1:CURR:LOW commands covered in
. The pulse width and delay are set with the
:SOUR1:PULS:WIDT and :SOUR1:PULS:DEL commands (see
).
Select source range
RANGe <n>
:SOURce[1]:CURRent:RANGe <n>
Select range for laser diode current source
Parameters
:<n> =
0 to 5.0
Specify source level (amps)
DEFault
500mA range
MINimum
500mA range
MAXimum
5A range
UP
Select next higher range
DOWN
Select next lower range
Query
:RANGe?
Query range for specified source
:RANGe? DEFault
Query *RST default source range
:RANGe? MINimum
Query lowest source range
:RANGe? MAXimum
Query highest source range
Description
This command is used to select the range for the laser diode current
source. Range is selected by specifying the approximate current magni-
tude that you will be using. The instrument will then go to the lowest
range that can accommodate that level.
Содержание 2520
Страница 250: ...A Specifications ...
Страница 254: ...B StatusandErrorMessages ...
Страница 262: ...C DataFlow ...
Страница 266: ...D IEEE 488BusOverview ...
Страница 281: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 285: ...F MeasurementConsiderations ...
Страница 308: ...G GPIB488 1Protocol ...
Страница 312: ...H ExamplePrograms ...
Страница 317: ...I ContinuousPulseMode ...