Model 2520 User’s Manual
Remote Operations
10-9
GPIB status indicators
The REM (remote), TALK (talk), LSTN (listen), and SRQ (service request) annunciators
show the GPIB bus status. Each of these indicators is described below.
REM
This indicator shows when the instrument is in the remote state. REM does not necessarily
indicate the state of the bus REN line, as the instrument must be addressed to listen with
REN true before the REM indicator turns on. When the instrument is in remote, all front
panel keys, except for the LOCAL key, are locked out. When REM is turned off, the
instrument is in the local state, and front panel operation is restored.
NOTE
If LLO is in effect, LOCAL will be locked out. OUTPUT ON/OFF is still opera-
tional in remote. If TRIG:SOUR is set to manual, the TRIG key will be active in
remote.
TALK
This indicator is on when the instrument is in the talker active state. Place the unit in the
talk state by addressing it to talk with the correct MTA (My Talk Address) command.
TALK is off when the unit is in the talker idle state. Place the unit in the talker idle state by
sending a UNT (Untalk) command, addressing it to listen, or sending the IFC (Interface
Clear) command.
LSTN
This indicator is on when the Model 2520 is in the listener active state, which is activated
by addressing the instrument to listen with the correct MLA (My Listen Address) com-
mand. LSTN is off when the unit is in the listener idle state. Place the unit in the listener
idle state by sending UNL (Unlisten), addressing it to talk, or sending IFC (Interface
Clear) command over the bus.
SRQ
You can program the instrument to generate a service request (SRQ) when one or more
errors or conditions occur. When this indicator is on, a service request has been generated.
This indicator stays on until the serial poll byte is read or all the conditions that caused
SRQ have been cleared. See
LOCAL key
The LOCAL key cancels the remote state and restores local operation of the instrument.
Pressing the LOCAL key also turns off the REM indicator and returns the display to nor-
mal if a user-defined message was displayed.
If the LLO (Local Lockout) command is in effect, the LOCAL key is also inoperative.
For safety reasons, the OUTPUT key can be used to turn the output off while in LLO.
Содержание 2520
Страница 250: ...A Specifications ...
Страница 254: ...B StatusandErrorMessages ...
Страница 262: ...C DataFlow ...
Страница 266: ...D IEEE 488BusOverview ...
Страница 281: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 285: ...F MeasurementConsiderations ...
Страница 308: ...G GPIB488 1Protocol ...
Страница 312: ...H ExamplePrograms ...
Страница 317: ...I ContinuousPulseMode ...