Model 2520 User’s Manual
Measurement Considerations
F-21
this purpose, measuring instruments should be placed on their lowest ranges. The configu-
ration that results in the lowest noise signal is the one that should be used.
Figure F-18
Eliminating ground loops
Figure F-19
Power line ground loops
Light
Some components, such as semiconductor junctions and MOS capacitors on semiconduc-
tor wafers, are excellent light detectors. Consequently, these components must be tested in
a light-free environment. While many test fixtures provide adequate light protection, oth-
ers may allow sufficient light penetration to affect the test results. Areas to check for light
leaks include doors and door hinges, tubing entry points, and connectors or connector
panels.
Instrument 1
Instrument 2
Instrument 3
Signal Leads
Ground
Loop
Current
Power Line Ground
Instrument 1
Instrument 2
Instrument 3
Power Line Ground