14-22
SCPI Command Reference
Model 2520 User’s Manual
Data elements
ELEMents <item list>
:FORMat:ELEMents [SENSe] <item list>
Specify data elements for data string
Parameters
<item list> = CURRent[1]
Includes laser diode source value
CURRent2
Includes detector #1 current reading
CURRent3
Includes detector #2 current reading
VOLTage[1]
Includes laser diode voltage reading
VOLTage2
Includes detector #1source value
VOLTage3
Includes detector #2 source value
TIME
Includes timestamp
STATus
Includes status information
NOTE
Each item in the list must be separated by a comma (i.e., VOLT1, CURR2,
TIME).
Query
:ELEMents?
Query elements in data string
Description
This command is used to specify the elements to be included in the data
string in response to the following queries:
:FETCh?
:READ?
:MEASure?
You can specify from one to all eight elements. Each element in the list
must be separated by a comma (,). These elements (shown in
) are explained as follows:
NOTE
An overflow reading reads as +9.9E37.
CURRent[1]
— This element provides the laser diode current source
value.
CURRent2
— This element provides the detector #1 current reading. If
no current reading is available, the NAN (not a number) value of
+9.91e37 is used.
CURRent3
— This element provides the detector #2 current reading. If
no current reading is available, the NAN (not a number) value of
+9.91e37 is used.
VOLTage[1]
— This element provides the laser diode voltage reading. If
voltage reading is available, the NAN (not a number) value of +9.91e37
is used.
VOLTage2
— This element provides the detector #1 voltage source
value.
Содержание 2520
Страница 250: ...A Specifications ...
Страница 254: ...B StatusandErrorMessages ...
Страница 262: ...C DataFlow ...
Страница 266: ...D IEEE 488BusOverview ...
Страница 281: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 285: ...F MeasurementConsiderations ...
Страница 308: ...G GPIB488 1Protocol ...
Страница 312: ...H ExamplePrograms ...
Страница 317: ...I ContinuousPulseMode ...