Model 2520 User’s Manual
Sweep Operation
7-7
Performing a log staircase sweep
Step 1: Configure source and measure.
Configure the Model 2520 source and measure functions as follows:
1.
Select the current source range by pressing LASER I
L
then EDIT, then use the
RANGE keys.
2.
Press LASER V
L
, then choose the desired measurement range.
3.
Use the DETECTOR 1 V
B
and DETECTOR 2 V
B
keys, and EDIT keys to set the
photodiode bias voltages to the desired values.
4.
Set the current measurement ranges by pressing the DETECTOR 1 I
PD
,
DETECTOR 2 I
PD
, and RANGE keys.
Step 2: Configure sweep.
Configure the sweep as follows:
1.
Press CONFIG then SWEEP.
2.
Select TYPE then press ENTER.
3.
Select LOG then press ENTER.
4.
At the prompts, enter the desired START, STOP, and NO-OF-POINTS values.
5.
From the CONFIGURE SWEEPS menu, select SWEEP COUNT, press ENTER,
then choose FINITE or INFINITE as desired.
6.
Press EXIT to return to normal display.
Step 3: Turn outputs on.
Press the ON/OFF OUTPUT key to turn the outputs on (blue OUTPUT indicator turns
on). The Model 2520 will output the photodiode bias voltages.
Step 4: Run sweep.
To run the sweep, press the SWEEP key. After the sweep is completed, turn the output off
by pressing the ON/OFF OUTPUT key.
Step 5: Read buffer.
Use the RECALL key to access the readings stored in the buffer.
Содержание 2520
Страница 250: ...A Specifications ...
Страница 254: ...B StatusandErrorMessages ...
Страница 262: ...C DataFlow ...
Страница 266: ...D IEEE 488BusOverview ...
Страница 281: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 285: ...F MeasurementConsiderations ...
Страница 308: ...G GPIB488 1Protocol ...
Страница 312: ...H ExamplePrograms ...
Страница 317: ...I ContinuousPulseMode ...