Model 2520 User’s Manual
Status Structure
11-13
Operation event register
The used bits of the Operation Event Register (shown in
lows:
•
Bit B0, Calibrating (Cal)
— Set bit indicates that the Model 2520 is calibrating.
•
Bits B1 through B4
— Not used.
•
Bit B5, Waiting for Trigger Event (Trig)
— Set bit indicates that the Model 2520
is in the trigger layer waiting for a TLINK trigger event to occur.
•
Bits B6 through B9
— Not used.
•
Bit B10, Idle State (Idle)
— Set bit indicates the Model 2520 is in the idle state.
•
Bits B11 through B15
— Not used.
Figure 11-5
Operation event status
Cal
(B0)
Trig
(B5)
—
(B9 - B6)
Idle
(B10)
—
(B15 - B11)
Operation
Condition Register
Operation
Event Register
Operation Event
Enable Register
&
&
&
OR
To Operation
Summary Bit
(OSB) of Status
Byte Register
(Figure 11-3)
Stat:oper:cond ?
Cal
(B0)
Trig
(B5)
—
(B9 - B6)
Idle
(B10)
—
(B15 - B11)
Cal
(B0)
Trig
(B5)
—
(B9 - B6)
Idle
(B10)
—
(B15 - B11)
Idle = In Idle State
Stat:oper: ?
:stat:oper:enab <NRf>
:stat:oper:enab ?
—
(B4 - B1)
—
(B4 - B1)
—
(B4 - B1)
Trg = Waiting for Trigger Event
Cal = Calibrating
& = Logical AND
OR = Logical OR
Содержание 2520
Страница 250: ...A Specifications ...
Страница 254: ...B StatusandErrorMessages ...
Страница 262: ...C DataFlow ...
Страница 266: ...D IEEE 488BusOverview ...
Страница 281: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 285: ...F MeasurementConsiderations ...
Страница 308: ...G GPIB488 1Protocol ...
Страница 312: ...H ExamplePrograms ...
Страница 317: ...I ContinuousPulseMode ...