Model 2520 User’s Manual
Example Programs
H-5
List sweep program
The program listing below performs a list (custom) sweep as covered in
program sets up the following operating modes:
•
Source mode: list sweep
•
Source current range: 500mA
•
Current list points: 200mA, 100mA, 400mA, 300mA, 500mA
•
Pulse delay points: 7ms, 4ms, 2ms, 8ms, 1ms
•
Pulse width points: 10
µ
s, 50
µ
s, 35
µ
s, 20
µ
s, 60
µ
s
•
Detector bias voltages: 5V
' Laser diode custom sweep program.
' 2520 primary address = 25.
OPEN “IEEE” FOR OUTPUT AS #1 ' Open IEEE-488 output path.
OPEN “IEEE” FOR INPUT AS #2 ' Open IEEE-488 input path.
PRINT #1, “INTERM CRLF” ' Set input terminator.
PRINT #1, “OUTTERM LF” ' Set output terminator.
PRINT #1, “REMOTE 25” ' Put 2520 in remote.
PRINT #1, “OUTPUT 25;*RST” ' Restore GPIB defaults.
PRINT #1, “OUTPUT 25;:FORM:ELEM VOLT1,CURR2,CURR3” ' Voltage/current data.
PRINT #1, “OUTPUT 25;:SOUR1:CURR:MODE LIST” ' Custom sweep mode.
PRINT #1, “OUTPUT 25;:SOUR1:LIST:CURR 0.2,0.1,0.4,0.3,0.5” ' I list.
PRINT #1, “OUTPUT 25;:SOUR1:LIST:DEL 7e-3,4e-3,2e-3,8e-3,1e-3” ' D list.
PRINT #1, “OUTPUT 25;:SOUR1:LIST:WIDT 10e-6,50e-6,35e-6,20e-6,60e-6”
PRINT #1, “OUTPUT 25;:SOUR2:VOLT 5” ' Detector 1 bias = 5V.
PRINT #1, “OUTPUT 25;:SOUR3:VOLT 5” ' Detector 2 bias = 5V.
PRINT #1, “OUTPUT 25;:OUTP1 ON” ' Turn on outputs.
CLS
PRINT #1, “OUTPUT 25;:READ?” ' Trigger sweep and read data.
PRINT #1, “ENTER 25” ' Address 2520 to talk.
LINE INPUT #2, R$ ' Input sweep data.
PRINT “Custom sweep data: “; R$
PRINT #1, “OUTPUT 25;:OUTP1 OFF” ' Turn off outputs.
END
Содержание 2520
Страница 250: ...A Specifications ...
Страница 254: ...B StatusandErrorMessages ...
Страница 262: ...C DataFlow ...
Страница 266: ...D IEEE 488BusOverview ...
Страница 281: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 285: ...F MeasurementConsiderations ...
Страница 308: ...G GPIB488 1Protocol ...
Страница 312: ...H ExamplePrograms ...
Страница 317: ...I ContinuousPulseMode ...