Model 2520 User’s Manual
SCPI Command Reference
14-53
COUNt?
:SYSTem:ERRor:COUNt?
Return the number of errors
Description
After sending this command and addressing the Model 2520 to talk, a
decimal number will be sent to the computer. That is the number of mes-
sages in the Error Queue.
CODE[:NEXT]?
:SYSTem:ERRor:CODE[:NEXT]?
Read oldest error (code only)
Description
This command is identical to the [:NEXT]? command, except only the
code is returned. The message itself is not returned. The error is cleared
from the queue.
CODE:ALL?
SYSTem:ERRor:CODE:ALL?
Read all errors (codes only)
Description
This query command is identical to the :ALL? command, except only
the codes are returned. The actual messages are not returned. All errors
are cleared from the queue.
CLEar
:SYSTem:CLEar
Clear Error Queue
Description
This action command is used to clear the Error Queue of messages.
Simulate key presses
KEY
:SYSTem:KEY <NRf>
Simulate key-press
Parameters
<NRf> =
1
RANGE
▲
key
2
EDIT
▼
key
3
EDIT
key
4
MENU key
5
DETECTOR 1 I
PD
key
6
FILTER key
7
DELAY key
8
EDIT key
9
COMPL key
10
EDIT
key
11
EXIT key
12
DETECTOR 2 V
B
key
13
DIG OUT key
14
COMM key
▲
▲
Содержание 2520
Страница 250: ...A Specifications ...
Страница 254: ...B StatusandErrorMessages ...
Страница 262: ...C DataFlow ...
Страница 266: ...D IEEE 488BusOverview ...
Страница 281: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 285: ...F MeasurementConsiderations ...
Страница 308: ...G GPIB488 1Protocol ...
Страница 312: ...H ExamplePrograms ...
Страница 317: ...I ContinuousPulseMode ...