B-6
Status and Error Messages
Model 2520 User’s Manual
Number
Error message
Event
1
Status register
2
Bit
+601
+602
+603
+604
+605
+606
+700
+701
+702
Lost data errors:
Reading buffer data lost
GPIB address lost
Power-on state lost
Calibration data lost
Calibration dates lost
GPIB communication language lost
Communication errors:
Invalid system communication
ASCII only with RS-232
DSP communication timeout
EE
EE
EE
EE
EE
EE
EE
EE
EE
Standard Event
Standard Event
Standard Event
Standard Event
Standard Event
Standard Event
Standard Event
Standard Event
Standard Event
3
3
3
3
3
3
3
3
3
+802
+803
+810
+830
+900
Additional command execution
errors:
OUTPUT blocked by interlock
Not permitted with OUTPUT off
OUTPUT blocked by over temp
Invalid with INF TRIG:COUNT
Internal System Error
EE
EE
EE
EE
EE
Standard Event
Standard Event
Standard Event
Standard Event
Standard Event
4
4
4
4
3
1. EE = Error Event
SE = Status Event
SYS = System Error Event
2. Use following queries to read status registers:
Standard Event: *ESR?
Operation Event: STAT:OPER?
Measurement Event: STAT:MEAS?
Questionable Event: STAT:QUES?
Table B-1 (continued)
Status and error messages
Содержание 2520
Страница 250: ...A Specifications ...
Страница 254: ...B StatusandErrorMessages ...
Страница 262: ...C DataFlow ...
Страница 266: ...D IEEE 488BusOverview ...
Страница 281: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 285: ...F MeasurementConsiderations ...
Страница 308: ...G GPIB488 1Protocol ...
Страница 312: ...H ExamplePrograms ...
Страница 317: ...I ContinuousPulseMode ...