Model 2520 User’s Manual
SCPI Command Reference
14-47
Set amplitudes
[:IMMediate][:AMPLitude] <n>
:SOURce2:VOLTage[:LEVel][IMMediate][:AMPLitude] <n>
Set detector #1 source amplitude
:SOURce3:VOLTage[:LEVel][IMMediate][:AMPLitude] <n>
Set detector #2 source amplitude
Parameters
<n> =
-20 to 20
Set source amplitude (volts)
DEFault
0V
MINimum
-20V
MAXimum
20V
Query
:VOLTage?
Query programmed source amplitude
:VOLTage? DEFault
Query *RST default amplitude
:VOLTage? MINimum
Query lowest allowable amplitude
:VOLTage? MAXimum
Query highest allowable amplitude
Description
These commands are used to immediately update the amplitude of the
detector #1 and detector #2 voltage bias sources (if the outputs are on).
If the outputs are off when either of these commands is sent, the source
value will be updated to the most recently programmed value when the
outputs are turned on.
Note that the maximum output current for both voltage sources is
100mA.
SOURce4
The following commands are used to set the logic levels of the digital output lines, and to
control the number of bits.
Setting digital output
[:LEVel] <NRf> | <NDN>
:SOURce4:TTL[:LEVel] <NRf> | <NDN>
Set digital output pattern
Parameters
<NRf>
=
0 to 15
Decimal format
<NDN>
#Bx
Binary format: x = 0000 to 1111
#Hx
Hexadecimal format: x = 0 to F
#Qx
Octal format: x = 0 to 17
Query
:TTL?
Query digital output value
Содержание 2520
Страница 250: ...A Specifications ...
Страница 254: ...B StatusandErrorMessages ...
Страница 262: ...C DataFlow ...
Страница 266: ...D IEEE 488BusOverview ...
Страница 281: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 285: ...F MeasurementConsiderations ...
Страница 308: ...G GPIB488 1Protocol ...
Страница 312: ...H ExamplePrograms ...
Страница 317: ...I ContinuousPulseMode ...