Model 2520 User’s Manual
SCPI Command Reference
14-27
OUTPut subsystem
This subsystem is used to control the three source outputs and query the state of the inter-
lock. These commands are summarized in
.
Turn sources on or off
[:STATe] <b>
:OUTPut[1][:STATe] <b>
Turn all three sources on or off
Parameters
<b> =
0 or OFF
Turn sources off (standby)
1 or ON
Turn sources on (operate)
Query
:OUTPut[1]?
Query state of sources
Description
This command is used to turn all three source outputs on or off simulta-
neously. Turning the sources off places the Model 2520 in the idle state.
NOTE
The :SOURce[1]:CLEar command will also turn the sources off.
Interlock status
TRIPped?
:OUTPut[1]:INTerlock:TRIPped?
Description
This query command is used to determine if the two interlocks have
been tripped. The tripped condition (“1”) means that the sources can be
turned on (interlock line at logic low level). A “0” will be returned if the
sources cannot be turned on (interlock line at logic high level).
See
and
for details on the interlocks.
SENSe subsystem
The SENSe subsystem is used to configure and control the measurement channels of the
Model 2520. SENSe[1] controls the laser diode voltage measurement channel, :SENSe2
controls the detector #1 measurement channel, and :SENSe3 controls the detector #2 mea-
surement channel.
Some of the commands are global, where a single command affects all three channels. For
example, filter commands affect all three channels simultaneously. Other commands are
unique to a specific measurement channel. For example, you can program a unique range
setting for each channel.
The commands for this subsystem are summarized in
.
Содержание 2520
Страница 250: ...A Specifications ...
Страница 254: ...B StatusandErrorMessages ...
Страница 262: ...C DataFlow ...
Страница 266: ...D IEEE 488BusOverview ...
Страница 281: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 285: ...F MeasurementConsiderations ...
Страница 308: ...G GPIB488 1Protocol ...
Страница 312: ...H ExamplePrograms ...
Страница 317: ...I ContinuousPulseMode ...