14-6
SCPI Command Reference
Model 2520 User’s Manual
Table 14-3
FORMat command summary
Command
Description
Default
parameter SCPI
:FORMat
:SREGister <name>
:SREGister?
[:DATA] <type>[<,length>]
[:DATA]?
:BORDer <name>
:BORDer?
:ELEMents <item list>
Select data format for reading status event registers
(ASCii, HEXadecimal, OCTal, or BINary).
Query format for reading status event registers.
Specify data format (ASCii, REAL, 32, or SREal).
Query data format.
Specify byte order (NORMal or SWAPped).
Query byte order.
Specify FETCh? and READ? data elements
(CURRent[1], VOLTage[1], CURRent2, VOLTage2,
CURRent3, VOLTage3, TIME, and STATus).
ASC
ASC
Note
VOLT1,
CURR2,
CURR3
✓
✓
✓
✓
✓
:CALCulate <item list>
:CALCulate?
:TRACe <item list>
:TRACe?
:ELEMents?
:SOURce4 <name>
:SOURce4?
Specify CALCulate data elements (CALCulate,
TIME, and STATus).
Query CALCulate data elements.
Specify TRACe data elements (VOLTage[1],
CURRent2, CURRent3, and TIME).
Query TRACe data elements.
Query FETCh? and READ? data format elements.
Specify SOURce4 data format (ASCii, HEXadecimal,
OCTal, or BIN).
Query SOURce4 data format.
CALC
VOLT1,
CURR2,
CURR3
ASC
Note: *RST default is NORMal. :SYST:PRES default is SWAPPed.
Table 14-4
OUTPut command summary
Command
Description
Default
parameter
SCPI
:OUTPut[1]
[:STATe] <b>
[:STATe]?
:INTerlock
:TRIPped?
Path to control all three outputs.
Turn all three sources on or off.
Query state of sources.
Path to interlock status:
Interlock tripped?: 1 (yes) or 0 (no).
OFF
✓
✓
✓
Содержание 2520
Страница 250: ...A Specifications ...
Страница 254: ...B StatusandErrorMessages ...
Страница 262: ...C DataFlow ...
Страница 266: ...D IEEE 488BusOverview ...
Страница 281: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 285: ...F MeasurementConsiderations ...
Страница 308: ...G GPIB488 1Protocol ...
Страница 312: ...H ExamplePrograms ...
Страница 317: ...I ContinuousPulseMode ...