Appendix
7-4
96296081
EN /
BA
Table 202
Guidelines and manufacturer’s declaration – electromagnetic immunity
The KARL STORZ AIDA
®
WD
300
is intended for use in the electromagnetic environment specified below.
The user of the
KARL STORZ AIDA
®
WD
300 should make sure that they are used in such an environment.
Interference
immunity tests
IEC 60601-1-2:2007
test level
Conformity level
Electromagnetic
environment – guidelines
Electrostatic
discharge (ESD) acc.
to
IEC 61000-4-2
± 6 kV contact
discharge
± 8 kV air discharge
Complies
± 8 kV contact
discharge
± 2kV, ± 4kV, ± 8 kV,
± 15kV
air discharge
Floors should be made of
wood or concrete or covered
with ceramic tiles. If floors are
covered with synthetic material,
the relative humidity must be at
least 30%.
Rapid transient
electric interferences/
bursts acc. to
IEC 61000-4-4
± 2 kV for mains lines
± 1 kV for input and
output lines
Complies
± 2 kV for mains lines
± 1 kV for input and
output lines
The quality of the supply
voltage should correspond to
that of a typical business or
hospital environment.
Surges acc.
to IEC 61000-4-5
± 1 kV push-pull
voltage
± 2 kV common-mode
voltage
Complies
± 1 kV push-pull
voltage
± 2 kV common-
mode voltage
The quality of the supply
voltage should correspond to
that of a typical business or
hospital environment.
Voltage dips, short
interruptions
and fluctuations
of the supply voltage
acc. to IEC 61000-
4-11
<5% U
T
*
(>95% dip in U
T
)
for 1/2 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30& dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 seconds
Conforms
<5% U
T
*
(>95% dip in U
T
)
for 1/2 cycle
Conforms
40% U
T
(60% dip in U
T
)
for 5 cycles
Conforms
70% U
T
(30% dip in U
T
)
for 25 cycles
Conforms
<5% U
T
(>95% dip in U
T
)
for 5 seconds
The quality of the supply
voltage should correspond to
that of a typical business or
hospital environment.
If the user of the device
requests continued functioning
even when interruptions of
the power supply occur, it
is recommended that the
device be supplied from a
power supply that is free of
interruptions.
Magnetic field at the
supply frequency
(50/60 Hz)
acc. to IEC 61000-
4-8
3 A/m
Complies
30 A/m
The magnetic fields at the
mains frequency should be at
the values typical of business
or hospital environments.
* Note:
U
T
is the mains alternating voltage before applying the test levels.
Содержание AIDA WD 300
Страница 1: ...KARL STORZ AIDA Rel 1 4...
Страница 3: ...System description KARL STORZ AIDA Release 1 4...
Страница 10: ...Safety instructions 1 4 96296083EN BA...
Страница 22: ...System description 2 12 96296083EN BB...
Страница 32: ...Installation 3 10 96296083EN BA 17 Click on Finish and restart the system 18 Log on as an administrator...
Страница 78: ...Maintenance 5 4 96296083EN BA...
Страница 81: ......
Страница 83: ...Software description KARL STORZ AIDA Release 1 4...
Страница 164: ...Data backup and reorganization 3 12 96296082EN BA 5 Click on Start The data are deleted from the drive...
Страница 166: ...Behavior in exceptional situations 4 2 96296082EN BA...
Страница 168: ...Support from the program supplier manufacturer 5 2 96296082EN BA...
Страница 169: ......
Страница 171: ...Instruction manual KARL STORZ AIDA WD 300...
Страница 184: ...Safety instructions 2 4 96296081EN BA...
Страница 196: ...Maintenance 5 4 96296081EN BA...
Страница 209: ......