Safety instructions
2-2
96296081
EN /
BA
2
CaUTiOn:
Only the software and hardware components which have been tested and approved
by KARL STORZ and listed in chapter 2.1.1 of the system description 96296083 may be used. The
maximum permitted values at the inputs and outputs of the device must be observed. If components
are connected which exceed these values, this will void the warranty.
2. 3 Safety precautions at the installation site
1
nOTE:
Only those parts of the system which are approved as medical devices are suitable for use
within the patient environment.
3
Warning:
The electrical installation of the operating room in which the device is installed and
operated must comply with the applicable IEC standards.
3
Warning:
To avoid the risk of an electric shock, the device must only be connected to a power
supply with a protective conductor.
3
Warning:
The device must only be operated under the ambient conditions specified (see chapter 1.7).
3
Warning:
Keep the device out of reach of patients.
3
Warning:
The device is only voltage-free when the power plug has been disconnected.
3
Warning:
The device must be positioned so as to ensure that the power connection can simply be
disconnected at any time.
3
Warning:
The device must be installed so as to ensure sufficient ventilation.
2. 4 Safety measures during operation
3
Warning:
It is the user’s responsibility to make sure all devices connected to the system and the
complete system are safe and operate properly before using them. The device should not be used if
any damage is evident.
3
Warning:
Avoid allowing liquids to enter the device. Do not store liquids in the vicinity of the device.
If liquids penetrate the device despite the precautionary measures, sufficient time should be allowed for
evaporation.
3
Warning:
The display of image signals which do not originate from a KARL STORZ device may only
be used for information purposes.
3
Warning:
For safety reasons, do not simultaneously touch the device output sockets and the patient.
3
Warning:
If the volume of the loudspeaker or headset is too high, an unexpected audio signal could
startle the attending physician. Check volume setting before each procedure.
Содержание AIDA WD 300
Страница 1: ...KARL STORZ AIDA Rel 1 4...
Страница 3: ...System description KARL STORZ AIDA Release 1 4...
Страница 10: ...Safety instructions 1 4 96296083EN BA...
Страница 22: ...System description 2 12 96296083EN BB...
Страница 32: ...Installation 3 10 96296083EN BA 17 Click on Finish and restart the system 18 Log on as an administrator...
Страница 78: ...Maintenance 5 4 96296083EN BA...
Страница 81: ......
Страница 83: ...Software description KARL STORZ AIDA Release 1 4...
Страница 164: ...Data backup and reorganization 3 12 96296082EN BA 5 Click on Start The data are deleted from the drive...
Страница 166: ...Behavior in exceptional situations 4 2 96296082EN BA...
Страница 168: ...Support from the program supplier manufacturer 5 2 96296082EN BA...
Страница 169: ......
Страница 171: ...Instruction manual KARL STORZ AIDA WD 300...
Страница 184: ...Safety instructions 2 4 96296081EN BA...
Страница 196: ...Maintenance 5 4 96296081EN BA...
Страница 209: ......