Appendix
KARL STORZ AIDA
®
WD
300
7-1
7
Appendix
7. 1 Information on electromagnetic compatibility (EMC)
3
Warning:
Medical electrical devices are subject to special precautions regarding Electromagnetic
Compatibility (EMC). Observe the EMC instructions in this appendix during installation and
commissioning.
The KARL STORZ AIDA
®
WD
300complies with
IEC 60601-1-2:2007 [CISPR 11 Class A]
and
therefore meets the EMC requirements of the Medical Device Directive (MDD) 93/42/EEC.
1
nOTE:
The tables and guidelines that are included in this appendix provide information to the customer
or user that is essential in determining the suitability of the device or system for the electromagnetic
environment of use, and in managing the electromagnetic environment of use to permit the device or
system to perform its intended use without causing interferences in other devices or systems or non-
medical electrical devices. If this device does cause electromagnetic interference with other devices, the
user is encouraged to try to correct this by one or more of the following measures:
•
reorient or relocate the receiving device
•
increase the separation between the individual devices
•
connect the devices to different electrical circuits.
If you have any further questions, please contact your local representative or our service department.
3
Warning:
The use of portable and mobile RF equipment may have an impact on this or other pieces
of medical equipment.
3
Warning:
The use of cables other than those specified in the KARL STORZ instruction manual may
result in increased emissions or decreased immunity of the
KARL STORZ
AIDA
®
WD
300. The cables
listed below have been shown to comply with the requirements of
IEC 60601-1-2:2007
. When using
cables other than those specified here, it is the responsibility of the user to ensure that they comply with
IEC 60601-1-2:2007
.
3
Warning:
The connection of other devices to the system should be avoided as this could jeopardize
smooth running. If it is necessary to link the system with other devices, it must be established that this
is possible without any interference.
3
Warning:
The device should not be used adjacent to or stacked with other equipment. If adjacent
or stacked use is necessary, the device or system should be observed to verify normal operation in the
configuration in which it will be used.
3
Warning:
Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used not closer than 30 cm (12 inches) to any part of the
WD
300
including cables specified by the manufacturer. Otherwise, degradation of the performance of this
equipment could result.
Содержание AIDA WD 300
Страница 1: ...KARL STORZ AIDA Rel 1 4...
Страница 3: ...System description KARL STORZ AIDA Release 1 4...
Страница 10: ...Safety instructions 1 4 96296083EN BA...
Страница 22: ...System description 2 12 96296083EN BB...
Страница 32: ...Installation 3 10 96296083EN BA 17 Click on Finish and restart the system 18 Log on as an administrator...
Страница 78: ...Maintenance 5 4 96296083EN BA...
Страница 81: ......
Страница 83: ...Software description KARL STORZ AIDA Release 1 4...
Страница 164: ...Data backup and reorganization 3 12 96296082EN BA 5 Click on Start The data are deleted from the drive...
Страница 166: ...Behavior in exceptional situations 4 2 96296082EN BA...
Страница 168: ...Support from the program supplier manufacturer 5 2 96296082EN BA...
Страница 169: ......
Страница 171: ...Instruction manual KARL STORZ AIDA WD 300...
Страница 184: ...Safety instructions 2 4 96296081EN BA...
Страница 196: ...Maintenance 5 4 96296081EN BA...
Страница 209: ......