Installation
KARL STORZ AIDA
®
WD
300
4-1
4
Installation
4. 1 Safety instructions
3
Warning:
Installation and commissioning of the device must only be performed by authorized
qualified technicians from KARL STORZ SE & Co. KG who have been trained in electrical or electronic
engineering or by an authorized contractual partner.
3
Warning:
Combinations of medical devices are only assured to be safe if they are identified as such
in the respective instruction manuals or the intended use and interface specifications of the products
used in combination permit this.
3
Warning:
The device must be installed so
as to ensure sufficient ventilation. so as to ensure sufficient
ventilation.
3
Warning:
Connect the supplied potential equalization cable to the potential equalization connector. In
doing so, observe the applicable national regulations.
3
Warning:
The electrical installations in the room in which the device is installed and operated must
comply with the applicable IEC standards.
3
Warning:
The device must only be connected using the power cord supplied by KARL STORZ or a
comparable power cord with a national test seal.
3
Warning:
The use of freely accessible multiple socket outlets is not permitted.
3
Warning:
Multiple socket outlets must not be placed on the floor.
3
Warning:
Additional devices that are connected to electronic medical devices must conform to the
respective IEC or ISO standards (e.g., IEC 60950 for data processing equipment).
Furthermore, all
configurations shall comply with the standard requirements for medical systems (see IEC 60601-1-1 or
Clause 16 of the 3rd Ed. of IEC 60601-1, respectively).
Any person who connects additional equipment
to medical electrical equipment configures a medical system, and is therefore responsible for ensuring
that the system complies with the requirements for medical electrical systems. Please note that local
laws take priority over the requirements mentioned above. If in doubt, consult the technical service
department or your local representative.
4. 2 Unpacking
Check the delivery for missing items and possible damage. File any complaints with the manufacturer
or supplier immediately.
Keep the original packaging for later use. This may be useful if the device has to be transported.
Содержание AIDA WD 300
Страница 1: ...KARL STORZ AIDA Rel 1 4...
Страница 3: ...System description KARL STORZ AIDA Release 1 4...
Страница 10: ...Safety instructions 1 4 96296083EN BA...
Страница 22: ...System description 2 12 96296083EN BB...
Страница 32: ...Installation 3 10 96296083EN BA 17 Click on Finish and restart the system 18 Log on as an administrator...
Страница 78: ...Maintenance 5 4 96296083EN BA...
Страница 81: ......
Страница 83: ...Software description KARL STORZ AIDA Release 1 4...
Страница 164: ...Data backup and reorganization 3 12 96296082EN BA 5 Click on Start The data are deleted from the drive...
Страница 166: ...Behavior in exceptional situations 4 2 96296082EN BA...
Страница 168: ...Support from the program supplier manufacturer 5 2 96296082EN BA...
Страница 169: ......
Страница 171: ...Instruction manual KARL STORZ AIDA WD 300...
Страница 184: ...Safety instructions 2 4 96296081EN BA...
Страница 196: ...Maintenance 5 4 96296081EN BA...
Страница 209: ......