Maintenance
5-2
96296081
EN /
BA
5. 2. 4 Required reprocessing materials
•
Disposable cloth or ready-to-use disinfectant cloth
•
Dry, low lint cloth
5. 2. 5 Wipe-down disinfection
1
nOTE:
To prevent liquid entering the device, it should not be sprayed with the disinfectant.
1
nOTE:
In order to preserve the conductivity of the contacts, you should never allow disinfectants to
come into contact with the connections on the front and back of the device. When the disinfectant
dries, residues can form which impair conductivity.
1.
Use a disposable cloth moistened with disinfectant or a soaked ready-to-use disinfectant cloth.
2. Wipe-clean the exterior surfaces of the medical device.
3. Allow the disinfectant to take effect, observing the instructions of the disinfectant manufacturer.
4. At the end of the necessary exposure time, wipe the surfaces with a dry low lint cloth.
5. 3 Maintenance and repair
5. 3. 1 Maintenance plan
Interval
Action
Responsibilities
Annually
Check and inspect device
KARL STORZ service technician
Every 4 years
Replace BIOS battery
KARL STORZ service technician
5. 3. 2 Changing the BIOS battery
The BIOS battery is changed by KARL STORZ service staff.
5. 3. 3 Servicing
Defective device components must be serviced and repaired exclusively by persons authorized by
KARL STORZ; all repair work must employ original KARL STORZ parts only.
All service work must only be performed by KARL STORZ or by an authorized agent.
3
Warning
No changes may be made to device without the manufacturer’s approval.
Содержание AIDA WD 300
Страница 1: ...KARL STORZ AIDA Rel 1 4...
Страница 3: ...System description KARL STORZ AIDA Release 1 4...
Страница 10: ...Safety instructions 1 4 96296083EN BA...
Страница 22: ...System description 2 12 96296083EN BB...
Страница 32: ...Installation 3 10 96296083EN BA 17 Click on Finish and restart the system 18 Log on as an administrator...
Страница 78: ...Maintenance 5 4 96296083EN BA...
Страница 81: ......
Страница 83: ...Software description KARL STORZ AIDA Release 1 4...
Страница 164: ...Data backup and reorganization 3 12 96296082EN BA 5 Click on Start The data are deleted from the drive...
Страница 166: ...Behavior in exceptional situations 4 2 96296082EN BA...
Страница 168: ...Support from the program supplier manufacturer 5 2 96296082EN BA...
Страница 169: ......
Страница 171: ...Instruction manual KARL STORZ AIDA WD 300...
Страница 184: ...Safety instructions 2 4 96296081EN BA...
Страница 196: ...Maintenance 5 4 96296081EN BA...
Страница 209: ......