KARL STORZ AIDA
®
2-11
Description of the program functions
‘
DICOM
‘
menu
Set up the DICOM interface in this menu.
1
NOTE:
Consult with your facility’s IT department as regards setting up the DICOM interface.
Setting up DICOM parameters
1. Under
Client AET
, enter the AE title which was assigned to the KARL STORZ AIDA
®
system by
your IT department.
2. Under
Modality
, select a Modality Type.
ES
: Endoscopy
US
: Ultrasound
OT
: Other
3. Under
Char. set
, select the character set in collaboration with your IT department.
4. The port describes the channel on which the DICOM service of the AIDA application is running.
This is required for the Storage Commitment and in order to check the connection between the
DICOM server and the AIDA application.
5. Pressing the
Encryption
button will take you to a selection screen where you can choose
between two encryption protocols for data transfer; ISCL (Internet Security Consulting Language)
or TLS (Transport Layer Security)
If no encryption protocol is selected, data will be transferred without any encryption.
6. If a check mark is placed next to ‘Local Cert. file’, the local computer’s certificates for TLS have
been imported on this computer. These are imported by pressing the
Browser
button, and are
usually imported from a USB stick. The certificates are managed by the hospital’s IT department.
7. If a check mark is placed next to ‘Server Cert. file’, certificates have been loaded for the DICOM
server used by the application, also via the
Browser
function. These server certificates are
managed by the hospital’s IT department.
8. Leave the
LogLevel
as the factory setting.
1
NOTE:
The LogLevel can be adjusted for service purposes, but should normally remain in factory
settings.
Содержание AIDA WD 300
Страница 1: ...KARL STORZ AIDA Rel 1 4...
Страница 3: ...System description KARL STORZ AIDA Release 1 4...
Страница 10: ...Safety instructions 1 4 96296083EN BA...
Страница 22: ...System description 2 12 96296083EN BB...
Страница 32: ...Installation 3 10 96296083EN BA 17 Click on Finish and restart the system 18 Log on as an administrator...
Страница 78: ...Maintenance 5 4 96296083EN BA...
Страница 81: ......
Страница 83: ...Software description KARL STORZ AIDA Release 1 4...
Страница 164: ...Data backup and reorganization 3 12 96296082EN BA 5 Click on Start The data are deleted from the drive...
Страница 166: ...Behavior in exceptional situations 4 2 96296082EN BA...
Страница 168: ...Support from the program supplier manufacturer 5 2 96296082EN BA...
Страница 169: ......
Страница 171: ...Instruction manual KARL STORZ AIDA WD 300...
Страница 184: ...Safety instructions 2 4 96296081EN BA...
Страница 196: ...Maintenance 5 4 96296081EN BA...
Страница 209: ......