Maintenance
KARL STORZ AIDA
®
5-1
5
Maintenance
5. 1 Safety instructions
3
WARNING:
Before performing any cleaning or maintenance work on the central device and the
connected devices, disconnect these from the power supply line!
3
WARNING:
All work may only be performed by qualified technicians trained in electrical or electronic
engineering, in compliance with the relevant occupational, safety and accident prevention regulations!
Prior to carrying out any work on the device, disconnect the device from the mains! A safety test
according to VDE 0751 and IEC 62353 must be carried out following service work.
3
WARNING:
By making the enclosed technical information available, KARL STORZ does not authorize
any service or repair work by unauthorized service personnel. Any tampering with instruments or
devices or unauthorized servicing or repair work to the device will render the warranty void.
5. 2 Reprocessing
5. 2. 1 Legal requirements
3
WARNING:
When carrying out any work on contaminated medical devices, the guidelines of the
Employers’ Liability Insurance Association and equivalent organizations striving to ensure personal
safety must be observed.
2
CAUTION:
National laws and regulations must be observed.
5. 2. 2 Safety instructions
3
WARNING:
Risk of infection: Incorrectly reprocessed medical devices expose patients, users and
third parties to a risk of infection as well as the risk that the medical device may malfunction. Observe
the manual ‘Cleaning, Disinfection, Care, and Sterilization of KARL STORZ Instruments’ and the
accompanying documentation.
2
CAUTION:
When preparing and using solutions, follow the chemical manufacturer’s instructions,
paying close attention to proper concentration, exposure time and service life. Incorrect concentration
may result in damage. Bear in mind the microbiological range of action of the chemicals used.
1
NOTE:
The manual ‘Cleaning, Disinfection, Care and Sterilization of KARL STORZ Instruments’ can be
downloaded or requested by visiting www.karlstorz.com.
5. 2. 3 Required reprocessing agents
•
Disinfectants
1
NOTE:
Due to their protein-fixating effect and possible material incompatibility, alcohol-based agents
must not be used.
The chemical manufacturer’s specifications concerning material compatibility must be observed.
5. 2. 4 Required reprocessing materials
• Disposable cloth or ready-to-use disinfectant cloth
• Dry, low lint cloth
Содержание AIDA WD 300
Страница 1: ...KARL STORZ AIDA Rel 1 4...
Страница 3: ...System description KARL STORZ AIDA Release 1 4...
Страница 10: ...Safety instructions 1 4 96296083EN BA...
Страница 22: ...System description 2 12 96296083EN BB...
Страница 32: ...Installation 3 10 96296083EN BA 17 Click on Finish and restart the system 18 Log on as an administrator...
Страница 78: ...Maintenance 5 4 96296083EN BA...
Страница 81: ......
Страница 83: ...Software description KARL STORZ AIDA Release 1 4...
Страница 164: ...Data backup and reorganization 3 12 96296082EN BA 5 Click on Start The data are deleted from the drive...
Страница 166: ...Behavior in exceptional situations 4 2 96296082EN BA...
Страница 168: ...Support from the program supplier manufacturer 5 2 96296082EN BA...
Страница 169: ......
Страница 171: ...Instruction manual KARL STORZ AIDA WD 300...
Страница 184: ...Safety instructions 2 4 96296081EN BA...
Страница 196: ...Maintenance 5 4 96296081EN BA...
Страница 209: ......