7. Clean the video recorder, shutter release
(1) Use the soft dry cloth to wipe the video recorder, shutter release.
(2) If it is difficult to wipe away the blemish, clean with soft cloth dipped in moderate detergent and
then air-dry it.
11.2 Replace the fuse
Replace the fuse is to replace the power adapter.
11.3 Replacement of power supply cord
Before replacing the power supply cord, please contact Kaixin Company; replace the power supply
cord under the guidance of Kaixin Company. Please use the power supply cord provided by Kaixin
Company.
11.4 Use and maintenance for the rechargeable battery
11.4.1 Use and replacement for the rechargeable battery
The battery pack (model HYLB-1614) is charged within main unit or with charger provided by
KAIXIN Company. Service personnel or operator can replace the battery pack.
Charging in the main unit
1. Plug the output port of power adapter into the DC power input port on the left side of main unit
to charge. When charging, the battery charge indicator blinks green; when fully charged, the
indicator is green and not flashing.
2. The shutdown charging time is about 6 hours; over-charging or discharging will shorten the
battery life. The full charged battery can be used about 4 hours.
Charging in the charger
1. The output port of power adapter plugs into the input port of Charger to charge (see the battery
charging sketch map). The charge indicator is green and blinking state when charging; the
charge indicator is green and no blinking when fully charged.
2. The charging time is about 3 hours, over-charging or discharging will shorten the battery life;
the full charged battery can be used about 3.5 hours.
Charger
Battery pack
12V
DC power input port
Fig. the battery charging sketch map
Attention:
1.The fuse is inside the power adapter. Fuse shall be replaced by qualified service personnel
who get KAIXIN approval.
2.
Before replacing the fuse, please contact KAIXIN, replace the fuse under the guidance of
KAIXIN.
3. Before replacing the fuse, you must disconnect the mains supply.
4. Fuse Type: T3.15AH250VAC
Attention:
Battery charger shall meet the requirements of the IEC60601-1 standard.