Ultrasonic Diagnostic Instrument (Vet)
User’s Manual V1.01
- 23 -
Chapter Seven
General Measurement
7.1 Distance Measurement
1. In B, B/B mode, freeze desired image, press direction keys
to move cursor to “Measure”
position of display area;
2. Press
key, the measurement methods are showed in the lower left of the screen, press direction
keys
to choose “1.Distance”, press
key again, the cursor will show “+”;
3. Press direction keys or operate mouse to move the “+” mark to desired position, press
key or
click left mouse to set the “+” mark position as the starting point of the measurement;
4. Press direction keys or operate mouse to move the “+” mark to the end point of the measurement.
A lighted dotted line appears between the starting point and the end point as the dashed locus of the
measurement. The measured value is automatically displayed at the built-in mark“+: ----mm” on
the right side of the screen;
5. Press
key or click right mouse to exchange the starting point and end point of the measurement;
6. Press
key or click left mouse to finish the first measurement;
7. Press
key again, repeat the steps 3~6 to complete the multi-group data measurement;
8. Press
key to quit the measurement status;
9. Press direction keys
to choose “Clear” and press
key, or use the mouse to click “Clear”
to clear all marks and data;
10.Press
key or click the middle mouse to return the real-time status.
7.2 Circumference/Area/Volume Measurement
Circumference/area/volume measurement with ellipse method
1. In B, B/B mode, freeze desired image, press direction keys
to move cursor to “Measure”
position of display area;
2. Press
key, the measurement methods are showed in the lower left of the screen, press direction
keys
to choose “2.Ellipse”, press
key again, the cursor will show “+”;
3. Press direction keys or operate mouse to move the “+” mark to desired position, press
key or
click left mouse to set the “+” mark position as the starting point of the measurement;
4. Press direction keys or operate mouse to move the “+” mark to the end point of the measurement,
at the same time the elliptic curve appears;
5. Press
key, the “
- +
” mark appears on the lower right of the image and hold down
or
key to
change the minor axis of the ellipse, or directly scroll the middle mouse button to change the minor
axis of the ellipse so as to satisfy the test area. The measured values are displayed at the built-in
characters “C: ----mm, A: ----mm
2
, V: ----cm
3
” on the right part of the screen automatically;
6. Press
key again to quit the minor axis status; Press
key or click right mouse to exchange the
starting point and end point;
7. Press
key or click left mouse to finish the first measurement;