Ultrasonic Diagnostic Instrument (Vet)
User’s Manual V1.01
- 25 -
circumference and area values are automatically displayed at the built-in mark “C: ----mm, A:
0.0mm
2
” on the right part of the screen;
5. Press
key again, the cursor will show “+”, repeat steps 3, 4 to complete the multi-group data
measurement;
6. Press
key to quit the measurement status;
7. Press direction keys
to choose “Clear” and press
key, or use the mouse to click “Clear”
to clear all marks and data;
8. Press
key or click the middle mouse to return the real-time status.
7.3 Slope/Heart rate/Cycle Measurement
In B/M, M frozen status, measure slope/hear rate/cycle. The method is identical with distance
measurement.
Note: Cycle measured in ms (millisecond)
HR denotes heart rate measured in times/minute (times per minute)
EF denotes slope coefficient measured in mm/s (millimeter per second)
Attention:
The accuracy of software measurement: distance measurement ≤ 0.1mm;
area measurement ≤ 0.1mm
2
; volume measurement ≤ 0.01cm
3
; heart rate
measurement ≤ 1bpm; cycle measurement ≤ 1ms. Due to differences in images
obtained by each user in different times, the actual object for the accuracy of the
measurement may be greater than the above-mentioned values.