Ultrasonic Diagnostic Instrument (Vet)
User’s Manual V1.01
- 5 -
Chapter Four
Operation Condition
4.1 Power supply
Adapter ratings: 100-240V~, 1.2-0.6A, 50-60Hz
Adapter model: BJE01-40-001M
Output of adapter: DC12.8V 3.0A
Main device rating: DC12.8V 3.0A
Internal supply voltage: DC11.1V
±
10%
4.2 Operation Environment
Ambient temperature: 10℃-40℃
Relative humidity: 30%-75% (without condensation)
Atmospheric pressure: 700hPa-1060hPa.
Altitude: < 2000 m
Overvoltage: Overvoltage Category
Ⅱ
Pollution degree: 2
4.3 Storage and Transportation
Ambient temperature: -20℃-55℃
Relative humidity: 30%-93% (without condensation)
Atmospheric pressure: 700hPa-1060hPa
Danger:
Do not use this equipment where flammable gas (such as anesthetic gas, oxygen
or hydrogen) or flammable liquid (such as alcohol) are present. Failure to do so may result
in explosion.
Warning:
Avoid using this equipment with high-frequency electric knife, high-frequency
therapy equipment or defibrillators and other electronic devices, or may an electric shock
occur to the patient.
Attention:
System should be avoided using in following environments:
1. Moist
2. Rain
3. Thunderstorm weather
4. No ventilation
5. Close to heat source (e.g. heaters, microwave ovens, ovens, water heaters, etc.)
6. Direct sunlight
7. Dramatic temperature change
8. Poisonous gas
9. Corrosive gas
10.Strong shock
11. Strong electromagnetic field (e.g. MRI)
Attention:
The mains voltage is varies with different countries or regions.
Attention:
Using radio transmitting equipment nearby the system may interfere with the
normal operation of the system. Prohibited carry or use of devices that can generate radio
waves within the room installed this system, such as cell phones, radio transceivers and
wireless remote control toys.