Ultrasonic Diagnostic Instrument (Vet)
User’s Manual V1.01
- 24 -
8. Press
key again, repeat the steps from 3 to 7 to complete the multi-group data measurement;
9. Press
key to quit the measurement status;
10. Press direction keys
to choose “Clear” and press
key, or use the mouse to click “Clear”
to clear all marks and data;
11. Press
key or click the middle mouse to return the real-time status.
Circumference/area measurement with trace method
1. In B, B/B mode, freeze desired image, press direction keys
to move cursor to “Measure”
position of display area;
2. Press
key, the measurement methods are showed in the lower left of the screen, press direction
keys
to choose “3.Trace”, press
key again, the cursor will show “+”;
3. Press direction keys or operate mouse to move the “+” mark to desired position, press
key or
click left mouse to set the “+” mark position as the start of the measurement;
4. Press direction keys or operate mouse to move the “+” mark to the end point of the measurement.
At the same time, a locus appears in the direction of operation between the two measurement
marks. The measured circumference value is displayed automatically at the built-in mark “C:
----mm” on the right part of the screen. Press
key or click left mouse to display at the built-in
mark “A: 0.0mm
2
” the value of the measured area formed by measurement line enclosure;
5. Press
key again, repeat the steps from 3 to 4 to complete the multi-group data measurement;
6. Press
key to quit the measurement status;
7. Press direction keys
to choose “Clear” and press
key, or use the mouse to click “Clear”
to clear all marks and data;
8. Press
key or click the middle mouse to return the real-time status.
Circumference/area measurement with point method
1. In B, B/B mode, freeze desired image, press direction keys
to move cursor to “Measure”
position of display area;
2. Press
key, the measurement methods are showed in the lower left of the screen, press direction
keys
to choose “4.Point”, press
key again, the cursor will show “+”;
3. Press direction keys or operate mouse to move the “+” mark to desired position, press
key or
click left mouse to mark the first point, continue to press direction keys or operate mouse, move
the “+” mark to next desired position, press
key or click left mouse to mark the second point;
By analogy, to mark all desired points;
Note: The number of points must be within 8 to 32.
4. When completed the marking of desired points, press
key or click right mouse, it appears trace
on the screen (trace is automatically drawn along the order of marking point until close the
beginning point and end point), the measured circumference and area values are automatically
displayed at the built-in mark “C: ----mm, A: 0.0mm
2
” on the right part of the screen; If
continuously marks the thirty-two point, it directly appears trace on the screen, the measured