6.4.12 Local Zoom and Local additive color
In real time mode, press
key to enter main-menu, press direction keys
to move cursor to
“Local Zoom” in the display area, press
key, a box appears. Press direction keys or operate mouse to
move the box to the position to be enlarged, the selected image be enlarged; Press
key to quit local
zoom status.
In the color display, the image selected by above operation will be enlarged and added color.
6.4.13 Image Left/Right Reverse
In real-time B, B/B, B/M mode, press
key to enter main-menu, press direction keys
to
move cursor to “L/R” in the display area, and then press direction keys
to achieve image
horizontal reverse. Or click left/right mouse also to achieve image horizontal reverse. The image left/right
reverse is the change of probe scanning direction. The probe scanning direction is indicated by the arrow
on the upper left area of the image.
6.4.14 Image Up/down Reverse
In real-time mode, press
key to enter main-menu, press direction keys
to move cursor to
“U/D” and then press direction keys
or click left/right mouse to achieve image vertical reverse.
6.4.15 Color Selection
In real-time mode, press
key to enter main-menu, press direction keys
to move cursor to
“Color” in the display area. Press direction keys
to achieve the conversion of eight kinds of colors
(including one kind of black and white).
6.4.16 Image Freeze/Unfreeze
In real-time mode, press
key or click the middle mouse to freeze the image; in frozen status,
press
key or click the middle mouse to unfreeze the image.
6.5 Gain adjustment in A mode
In real-time A mode, press
key to enter main-menu, press direction keys
to move cursor to
“Gain” in the display area. Press direction key
to increase image gain, press direction key
to reduce it.
6.6 Puncture guide line
In real-time B mode, press
key to enter main-menu, press direction keys
to move cursor to
“Puncture” and then press direction keys
to choose line 1, press
key to confirm, two puncture
guide lines appear on the screen, press direction keys
to change the angle of the first puncture
guide line, press direction keys
to change the start position of the first puncture guide line;
Press
key, the cursor is located on the “Puncture”, press direction keys
to choose line 2,
press
key to confirm, press direction keys
to change the angle of the second puncture guide line,
press direction keys
to change the start position of the second puncture guide line. Press
key to
quit the puncture guide status.
6.7 Body Mark and Probe Mark
This product contains 27 body marks that are divided into two pages when display. The operation