Ultrasonic Diagnostic Instrument (Vet)
User’s Manual V1.01
- 6 -
Chapter Five
System Installation and Check
Warning:
To avoid the risk of electric shock, this equipment must only be connected to a
supply mains with protective earth.
Warning:
1. When instrument works abnormally, do stop working, turn off the power and check the
reason, then contacts the KAIXIN Company about it.
2. Turn off power and pull out of the plug from socket after each ultrasonic diagnostic
operation.
3. It is forbidden to drag and press the power and probe cables emphatically; regularly inspect
whether there is pull-apart and bareness, if there is the phenomena like this, turn off power
supply immediately, stop using it and change it for new one.
4. It is forbidden to load and unload the probe or move the instrument in galvanic to avoid
danger of safety.
5. Pull out of the plug from socket after operation in thunderstorm weather to avoid the
instrument being damaged by lightening.
6. If the temperature changes greatly in short time will cause vapor recovery inside of
instrument, the case may damage the instrument.
7. The instrument is turned off completely only by disconnecting the power supply from the
wall socket.
Warning:
1. All plugs of instruments of this system shall be connected into the power socket with
protectively earth on the wall and the socket must meet the requirement of power rating of
instrument. Use of multiple portable socket-outlets may affect protective earth to make
leakage currents exceed the safety requirements.
2. Please follow the correct electrical connections method to connect the power supply and
earth, otherwise there will be danger of electric shock. Do not connect the grounding wire
to any gas pipe or water pipe, or it may cause bad grounding and danger of explosion.
3. Prohibit the live parts of the equipment or other devices (such as various signal input and
output ports, etc.) contact with the patient, if this equipment or other equipment has
failure, the patient will have danger of electric shock.
4. Additional equipment connected to the medical electrical equipment must comply with the
respective IEC standards (e.g. IEC60950 for data processing equipment, IEC 60601-1 for
medical electrical systems ). Anybody connecting additional equipment to medical
electrical equipment configures a medical system and is therefore responsible that the
system complies with the requirements for medical electrical systems. Attention is drawn
to the fact that local laws take priority over the above mentioned requirements. When
more than one device is connected to use, the accumulation of leakage current may cause
the danger of security. If in doubt, consult your local representative or the technical service
department.
5. If the integrity of the external protective conductor in the installation or its arrangement is
in doubt, equipment shall be operated from its internal electrical power source.
Warning:
The waterproof grade of the equipment is IPX4 (no adverse effect on splashing
water in all directions). Splash any liquid on the equipment may damage it. If you
accidentally splash liquid on the equipment, please immediately turn off the power and
contact your service representative.