5.1 System placement

Please carefully read through and fully understand the safety cautions before moving and placing the

system.

1. Unpack the instrument case and check the goods for its completeness according to the packing list.

2. Place the instrument on a stable and leveled position.

3. Leave adequate space of 20 centimeters as minimum from rear, left and right side of the instrument.

4. Place the instrument on a desktop to use; adjust the angle of the support bracket to adjust the screen

visual angle.

5.2 Ultrasonic probe installation

Warning:

1. Do not use the probe not provided by our company, otherwise the equipment and the probe

will cause damage, and may cause fire in extreme cases.

2. Check the ultrasonic probe and connecting cable after diagnostic operation. Use of defective

probe may cause electric shock.

3. Do not strike the probe; using the damaged probe may cause electric shock to the patient.

4. Unauthorized disassembly of the probe shall be prohibited as it may cause electric shock.

Attention:

1. Turn off the ultrasonic system before disconnecting the probe. Disconnecting the probe with

system power on may damage the system or probe.

2. Before disconnecting the probe, place the cable and probe on a stable and leveled position so

that the probe may not be damaged or injury person by unexpected fall.

3. Freeze the instrument when instrument is start-up without operation to increase of service

life of probe.

4. Repeat available machine time should be more than 5 minutes to avoid turn on/off power

supply in short time.

Danger:

Use together with flammable anaesthetic, it may result in explosion.

Attention:

Probe is highly sensitive to shake, be used with caution. About probe’s use and

cleaning, the details see the relevant sections.

Warning:

The power adapter, probe, shutter release, battery and power supply cord as

described in this section are replaced by operator. But these parts must be provided by

KAIXIN or his authorized supplier.

Attention:

Adequate space from rear, left and right side of the machine shall be reserved,

or the machine may malfunction under excessive heat inside the enclosure.

Содержание K10

Страница 1: ...KAIXIN K10 Ultrasonic Diagnostic Instrument Vet User s Manual Xuzhou Kaixin Electronic Instrument Co Ltd ...

Страница 2: ...ess written permission of Kaixin Statement Kaixin has the final explanation right of this user s manual Kaixin was considered responsible for the safety reliability and performance in case of meeting all the following requirements 1 Assembly expansion readjustment improve and repair are all performed by professionals recognized by Kaixin 2 All replacement parts and accessories consumables involved...

Страница 3: ...ir the system by non Kaixin designated personnel 5 This product is intended to provide clinical diagnostic data for the doctor The doctor shall be responsible for the diagnostic process Kaixin shall not be liable for any problems arising out of the process 6 Be sure to back up important data to external storage media such as notebooks 7 Due to operator s error or abnormal condition causing the dat...

Страница 4: ...nexpected use condition that may result in light injury or property loss or affecting the use if not avoided 2 Safety Symbols Symbols Meaning Symbols Meaning Type B applied part Up Power switch Keep dry Direct current Fragile Power supply indication Stacking limit by number Adapter connection indication Temperature limits Storage and transport Battery charge indicator Humidity limitation Storage a...

Страница 5: ... check before and after operation 9 5 10 System reset 10 Chapter Six Functional Operation 11 6 1 Startup and Shutdown 11 6 2 System functions setting 11 6 2 1 Time setting 11 6 2 2 TV Mode setting 11 6 2 3 Key Sound setting 11 6 2 4 Screensaver setting 11 6 2 5 Image Filter setting 12 6 2 6 Grid setting 12 6 2 7 Wifi setting 12 6 2 8 Background color setting 12 6 2 9 User name setting 13 6 2 10 Ch...

Страница 6: ...view check list 20 6 12 2 Delete check lists 21 6 12 3 Transfer check lists to workstation 21 6 13 Obstetric list management 21 6 13 1 Save and view obstetric list 21 6 13 2 Delete obstetric lists 22 6 13 3 Transfer obstetric lists to workstation 22 Chapter Seven General Measurement 23 7 1 Distance Measurement 23 7 2 Circumference Area Volume Measurement 23 7 3 Slope Heart rate Cycle Measurement 2...

Страница 7: ... 1 Preparation before use 36 10 5 2 Start using PC side network workstation software 37 10 6 Change the language for ultrasound workstation software 37 10 6 1 Change the language for mobile and PC side software kx_station_v 37 10 6 2 Change the language for PC side network workstation software Network Printer 37 Chapter Eleven System Maintenance 38 11 1 Maintenance by users 38 11 1 1 System cleani...

Страница 8: ...ame correlation function 6 Image post process function 7 Edge enhancement function 8 Single point multi point focusing function 9 Local zooming function 10 Image freeze unfreeze 11 Vertical horizontal reverse 12 Color display function 13 Puncture guide 14 Image filter function 15 Save image by one key 16 Ultrasound workstation manages image and data 17 Case information image annotation and auto ti...

Страница 9: ...2 1 Structure composition of the instrument K10 ultrasonic diagnostic instrument Vet is composed of main unit and probe etc 2 2 Components name Figure K10 main unit sketch map 2 3 Parts of the probe Take 6 5MHz micro convex probe for example Figure Parts name of 6 5MHz micro convex probe ...

Страница 10: ... surface when performing ultrasonic diagnosis 2 Cable To connect the probe to the probe connector 3 Probe connector To connect the probe to ultrasonic diagnostic instrument 4 Lock loosen knob To lock loosen the connection of probe to ultrasonic diagnostic instrument 2 4 Function keys description SN Function keys Real time status function Freeze status function 1 Mode Selection Text input Transfer ...

Страница 11: ...y probe 1 PC 3 Power adapter 1 PC 4 PC or mobile ultrasound workstation software 1 PC 3 2 Optional parts 1 3 5 MHz convex array probe 2 3 5 MHz linear array loin probe 3 4 0 MHz convex array transrectal probe 4 6 5MHz animal transrectal linear array probe 5 7 5 MHz high frequency linear array probe 6 Video recorder 7 Shutter release 8 Probe holder 9 Mouse 10 Charger ...

Страница 12: ... are present Failure to do so may result in explosion Warning Avoid using this equipment with high frequency electric knife high frequency therapy equipment or defibrillators and other electronic devices or may an electric shock occur to the patient Attention System should be avoided using in following environments 1 Moist 2 Rain 3 Thunderstorm weather 4 No ventilation 5 Close to heat source e g h...

Страница 13: ...safety requirements 2 Please follow the correct electrical connections method to connect the power supply and earth otherwise there will be danger of electric shock Do not connect the grounding wire to any gas pipe or water pipe or it may cause bad grounding and danger of explosion 3 Prohibit the live parts of the equipment or other devices such as various signal input and output ports etc contact...

Страница 14: ...isassembly of the probe shall be prohibited as it may cause electric shock Attention 1 Turn off the ultrasonic system before disconnecting the probe Disconnecting the probe with system power on may damage the system or probe 2 Before disconnecting the probe place the cable and probe on a stable and leveled position so that the probe may not be damaged or injury person by unexpected fall 3 Freeze t...

Страница 15: ... the system push the Battery cover buckle on the battery cover on the back of main unit remove the battery cover pull the strap on the battery and take out the battery 5 4 Shutter release installation Install shutter release to foot switch port as shown in the figure 5 5 Video recorder installation 1 Turn off the system connect the equipotential terminal of the video recorder to the earthing 2 Con...

Страница 16: ...orking condition and free of abnormity such as cracks and loosened screws 5 No damage on cables power cable etc and hard up on its connection 6 Check probe and its connections to ensure they are free of abnormity such as scuffing drop off or contamination If the contaminant is present clean disinfect the contaminated objects as specified in relevant sections 7 No barriers around the intake of equi...

Страница 17: ...on 6 Check the instrument to ensure that the phenomenon of local high temperature will not appear 5 10 System reset In case of abnormal screen display or no working for system operation try to restart the system by turning on off the main unit power Attention If the overheat acoustic lens is placed on the patient s skin heat injury may occur Attention Thoroughly clean the coupling gel on the probe...

Страница 18: ... setting interface 6 2 2 TV Mode setting 1 In real time mode press key to enter main menu press direction keys to select Preset 2 Press key to enter setting interface 3 Press direction keys to select Mode 4 Press direction keys to achieve TV mode conversion between PAL and NTSC 5 Press key to confirm this setting and quit setting interface 6 2 3 Key Sound setting 1 In real time mode press key to e...

Страница 19: ...mm 02 shows line grid 03 shows dot grid 04 shows 10mm spacing dot like scale 05 shows 2mm spacing dot like scale 06 show 10mm spacing and 2mm spacing dot like scales 07 shows longitudinal 1mm spacing grid When the image depth is less than 10cm 08 shows longitudinal 5mm spacing grid 09 shows longitudinal 10mm spacing grid 5 Press key to confirm this setting and quit setting interface Note Grid is d...

Страница 20: ...o point to character or press key move the cursor to the position need to deleted character and then press directions keys to move cursor to character press key delete input content and retype 7 Press key to quit the character input menu press key again to quit the cursor status press key to return the frozen status 6 2 10 Chinese English setting 1 In real time mode press key to enter main menu pr...

Страница 21: ... mouse to increase the image parameter value 6 4 1 BackLight brightness and contrast adjustment The brightness and contrast level of the screen is one of the most important factors for the image quality The brightness and contrast adjustment should be done in relation to the ambient brightness Therefore the actual brightness and contrast level shall be adjusted according to the specific environmen...

Страница 22: ...pth range selection 6 4 7 Frequency Adjustment Frequency conversion In real time mode press key to enter main menu press direction keys to move cursor to Freq in the display area Press direction keys to achieve frequency conversion 6 4 8 Frame Correlation Adjustment In real time B B B mode press key to enter main menu press direction keys to move cursor to Frame Avg in the display area Press direc...

Страница 23: ...ress direction keys to move cursor to Color in the display area Press direction keys to achieve the conversion of eight kinds of colors including one kind of black and white 6 4 16 Image Freeze Unfreeze In real time mode press key or click the middle mouse to freeze the image in frozen status press key or click the middle mouse to unfreeze the image 6 5 Gain adjustment in A mode In real time A mod...

Страница 24: ...ng images with one key you should always pay attention to the remaining frames within the inter memory of main unit to prevent invalid storage After 420 images have been saved delete it manually 6 9 Image management 6 9 1 Save the image Storage to main unit 1 In B B B B M M mode freeze the image 2 Press key a Save prompt appears on lower right corner of the screen 3 Press direction keys or click l...

Страница 25: ... in the U disk such as choose 004 5 Press key to read out the image stored in the frame 004 Img character showed at the lower right corner of the screen 6 Press key to quit saving status and press key to return to real time status Explanation 1 When reading images in the main unit you must choose the image code with reading images in the U disk must choose the image code with 2 When the U disk is ...

Страница 26: ...e 004 is transferred to ultrasound workstation 3 Press key to quit saving or reading image status 4 Press key to return to real time mode Note The specific functions of ultrasound workstation see Chapter Ten Ultrasound workstation software instructions 6 9 6 Batch images transfer to workstation 1 In B B B B M M mode freeze the image press key to transfer all the stored images to the ultrasound wor...

Страница 27: ...o Caps and then press key or click left mouse to achieve capital and small letter conversion If the cursor point to Shift press key or click left mouse to achieve the conversion between the letter and punctuation 3 Press direction keys to choose numbers or characters and press key to confirm or operate mouse to click numbers or characters to input 4 After inputting ID press key the screen below sh...

Страница 28: ...e cursor to Checklist and press key or use the left mouse key to click Checklist to enter check list interface 3 Press key a symbol prompt appears at the top right of the screen after the prompt disappear all the check lists will be transferred to ultrasound workstation 4 Press key or click right mouse to exit check list interface press key or click the middle mouse to return to real time status N...

Страница 29: ...tus 6 13 3 Transfer obstetric lists to workstation 1 In the B B B mode freeze the desired image 2 Press direction keys to move cursor to OB List press key or use the left mouse key to click OB List to enter obstetric list interface 3 Press key a symbol prompt appears at the top right of the screen after the prompt disappear all the obstetric lists will be transferred to ultrasound workstation 4 Pr...

Страница 30: ...o choose Clear and press key or use the mouse to click Clear to clear all marks and data 10 Press key or click the middle mouse to return the real time status 7 2 Circumference Area Volume Measurement Circumference area volume measurement with ellipse method 1 In B B B mode freeze desired image press direction keys to move cursor to Measure position of display area 2 Press key the measurement meth...

Страница 31: ... by measurement line enclosure 5 Press key again repeat the steps from 3 to 4 to complete the multi group data measurement 6 Press key to quit the measurement status 7 Press direction keys to choose Clear and press key or use the mouse to click Clear to clear all marks and data 8 Press key or click the middle mouse to return the real time status Circumference area measurement with point method 1 I...

Страница 32: ...return the real time status 7 3 Slope Heart rate Cycle Measurement In B M M frozen status measure slope hear rate cycle The method is identical with distance measurement Note Cycle measured in ms millisecond HR denotes heart rate measured in times minute times per minute EF denotes slope coefficient measured in mm s millimeter per second Attention The accuracy of software measurement distance meas...

Страница 33: ... click left mouse to set the mark position as the starting point of the measurement 5 Press direction keys or operate mouse to move the mark to the end point of the measurement At the same time a lighted dotted line appears between the start and the end as the dashed locus of the measurement The measured value is automatically displayed at the built in mark mm on the right side of the screen G Aan...

Страница 34: ...n to select Weight range1 300kg press key or click left mouse again to confirm 5 Measure the backfat thickness with distance measurement method the backfat thickness is displayed in real time in the right area of the screen measure loin with distance measurement method the loin is displayed in real time in the right area of the screen press key or click left mouse to finish the loin measurement th...

Страница 35: ... desired measurement site the trailing edge of shoulder P1 the tenth rib P2 the lumbosacral junction P3 2 Put ultrasonic coupling gel to the appropriate site of swine s back in order to make good contact between swine skin and sound window The probe is placed against the skin of the animal and is kept perpendicular to the back move the probe gently make sure no air bubbles between the skin and pro...

Страница 36: ...probe Keep the swine quiet during the test you can feed the swine during the test 3 After the image is automatically frozen the backfat thickness and loin thickness are measured and the lean percentage is automatically displayed in the right area of the screen 4 Press key or click the middle mouse to return the real time status Explanation Before measuring the swine s lean percentage you must ente...

Страница 37: ...ssion 3 After finish the operation you can use it Get PC side ultrasound workstation software kx_station_v According to the information provided in the card in the package get KAIXIN ultrasound workstation software copy to the computer open the workstation software folder double click the icon you can use Get PC side network workstation software Network Printer According to the information provide...

Страница 38: ...rasound workstation software Note 1 You first turn on the Wifi of K10 equipment and then set up the phone s WLAN After WLAN connection is successful open the ultrasound workstation software 2 The Wifi network ID of K10 equipment is the default number displayed on the nameplate of K10 such as 1808880 3 When the network is disconnected the wireless signal icon on the K10 screen disappears 10 3 2 Sta...

Страница 39: ...For example Local Internal Storage kxStation image DSC007_20180312164847 bmp 10 3 2 2 Check lists management 1 After the connection between K10 equipment and kx_station_v is successful the lower left corner of K10 screen shows the wireless signal icon 2 The check lists stored inside of K10 equipment will be transferred to the ultrasound workstation a In the B B B mode freeze the desired image b Pr...

Страница 40: ...ck list interface click import click the database to import the saved database into the current check list Generate report In the check list interface click report the check list is generated as a report Empty the check list In the check list interface click empty to delete all check lists 10 3 2 3 Obstetric lists management 1 After the connection between K10 equipment and kx_station_v is successf...

Страница 41: ...ord can query For example Note Whether which methods to inquire you must select the checkbox in front of the search field 2 After setting the search conditions tap to get the obstetric list of queries tap to give up the inquire 3 According to the query conditions to get the desired result then tap inquire to enter the query dialog box the checkbox is not selected click default display all lists Im...

Страница 42: ...0 3 2 6 Quit workstation software Tap quit to quit mobile ultrasound workstation software 10 4 PC side ultrasound workstation software kx_station_v The usages for PC side ultrasonic workstation software are similar to mobile workstation it communicates with K10 equipment via Wifi Wi Fi 10 4 1 Preparation before use 1 Install the PC side ultrasonic workstation software kx_station_v on the computer ...

Страница 43: ... with K10 equipment via a network cable It can transfer the contents of the screen to the network workstation The specific connection method is shown in the figure below 10 5 1 Preparation before use 1 Turn on the K10 equipment 2 Use a network cable to connect the network port of ultrasonic diagnostic instrument and the computer s network port and then installs the printer on your computer 3 Unzip...

Страница 44: ...Printer sees its instructions for use 10 6 Change the language for ultrasound workstation software 10 6 1 Change the language for mobile and PC side software kx_station_v 1 Go to Settings of the phone or Pad tap Language input tap Language and choose the used language 2 After finish the settings enter the ultrasound workstation again the ultrasound workstation interface appears the changed languag...

Страница 45: ...fer to the instructions of the solution for details Non glutaraldehyde based disinfectant Chemical Name Reagent Name Step Phthalaldehyde solution 0 55 Cidex OPA Please refer to the instructions of the solution for details Please follow the instructions about disinfectant concentration and disinfection method as well as the precautions about disinfectants provided by disinfectant provider But do no...

Страница 46: ... window such as scuffing cracks peeling bulge and drop off 4 Clean the probe cable and its connector 1 Clean the probe cable and its connector with soft dry cloth 2 In case of die hard blots clean with soft cloth dipped in moderate detergent and then air dry it 5 Clean the LED screen Use a soft cloth dipped in glass cleaner to clean the LED screen and then air dried 6 Clean the control panel shell...

Страница 47: ...ain unit to charge When charging the battery charge indicator blinks green when fully charged the indicator is green and not flashing 2 The shutdown charging time is about 6 hours over charging or discharging will shorten the battery life The full charged battery can be used about 4 hours Charging in the charger 1 The output port of power adapter plugs into the input port of Charger to charge see ...

Страница 48: ...ion or fire 9 Do not disassemble the battery in any way which will lead to the leakage explosion or fire 10 Do not charge the battery near the heat source or extra hot environment which will lead to the leakage explosion or fire 11 Do not put the battery into the microwave oven or pressure vessel which will lead to the leakage explosion or fire 12 Do not mixed use the battery together with one off...

Страница 49: ...lug socket of the instrument or probe connectors They shall be properly contacted Image display is not clear 1 Adjust the total gain near field gain far field gain 2 Adjust the backlight brightness and contrast level Control panel malfunction Restart the system by turning off the main unit power 11 6 Periodic Safety Checks To ensure the system performance and safety it must be checked after using ...

Страница 50: ...Shockproof measures have been taken inside the packing case to allow for transport by air railway land and sea The goods shall not be exposed to poor weather conditions like rain and snow nor shall the goods be placed upside down bumped knocked or over stacked Chapter Thirteen Safety Classification 1 Classified according to electric shock protection type Class I internally powered equipment 2 Clas...

Страница 51: ...ixin Mansion C 01 Economic Development Zone Xuzhou Jiangsu China Post Code 221004 Tel 86 516 87732932 87733758 Fax 86 516 87732932 87792848 Website www kxele com E mail info kxele com Information contained in this manual is subject to change without further notice ...