Ultrasonic Diagnostic Instrument (Vet)
User’s Manual V1.01
- 7 -
5.1 System placement
Please carefully read through and fully understand the safety cautions before moving and placing the
system.
1. Unpack the instrument case and check the goods for its completeness according to the packing list.
2. Place the instrument on a stable and leveled position.
3. Leave adequate space of 20 centimeters as minimum from rear, left and right side of the instrument.
4. Place the instrument on a desktop to use; adjust the angle of the support bracket to adjust the screen
visual angle.
5.2 Ultrasonic probe installation
Warning:
1. Do not use the probe not provided by our company, otherwise the equipment and the probe
will cause damage, and may cause fire in extreme cases.
2. Check the ultrasonic probe and connecting cable after diagnostic operation. Use of defective
probe may cause electric shock.
3. Do not strike the probe; using the damaged probe may cause electric shock to the patient.
4. Unauthorized disassembly of the probe shall be prohibited as it may cause electric shock.
Attention:
1. Turn off the ultrasonic system before disconnecting the probe. Disconnecting the probe with
system power on may damage the system or probe.
2. Before disconnecting the probe, place the cable and probe on a stable and leveled position so
that the probe may not be damaged or injury person by unexpected fall.
3. Freeze the instrument when instrument is start-up without operation to increase of service
life of probe.
4. Repeat available machine time should be more than 5 minutes to avoid turn on/off power
supply in short time.
Danger:
Use together with flammable anaesthetic, it may result in explosion.
Attention:
Probe is highly sensitive to shake, be used with caution. About probe’s use and
cleaning, the details see the relevant sections.
Warning:
The power adapter, probe, shutter release, battery and power supply cord as
described in this section are replaced by operator. But these parts must be provided by
KAIXIN or his authorized supplier.
Attention:
Adequate space from rear, left and right side of the machine shall be reserved,
or the machine may malfunction under excessive heat inside the enclosure.