(78 mm)
(61 mm)
(265 mm)
(
B
3.
6
m
m
)
Small Battery Drive II User Manual DePuy Synthes 35
Warnings:
The Small Battery Drive II batteries must
not be cleaned following the Mechanical / Automated
Cleaning Instructions with Manual Pre-Cleaning.
Important:
•
Manual pre-cleaning prior to automated
cleaning / disinfection is important to ensure that
cannulations and other difficult to access areas
are clean.
•
Alternative cleaning / disinfection procedures
other than in the procedure described below
(including manual precleaning) have not been
validated by Depuy Synthes.
1.
Remove debris.
Rinse the device under running cold
tap water for a minimum of 2 minutes. Use a soft
lint-free cloth or soft-bristled brush to assist in
removing gross soil. For cannulations of the
handpiece and attachments, the cleaning brush
(519.400) shown below should be used.
Notes:
•
Brushes and other cleaning tools shall be either
single use items or, if reusable, be decontaminated
at least daily using a solution as detailed in section
“3. Spray and wipe”.
•
Brushes shall be inspected before daily use and
discarded if they have degraded to the point where
they may scratch instrument surfaces or be
ineffective due to worn or missing bristles.
Warnings:
•
Do not immerse the handpiece, batteries, battery
casings or attachments in aqueous solutions or in
an ultrasonic bath.
•
Do not use pressurized water as this will cause
damage to the system.
Precaution:
Do not use pointed objects for cleaning.
2.
Manipulate moving parts.
Manipulate all moving
parts such as the triggers, sleeves and switches under
running tap water to loosen and remove gross debris.
Cleaning brush (519.400)
Care and Maintenance
Cleaning and Disinfection
Automated Cleaning Instructions
with Manual Pre-Cleaning
Содержание DePuy Synthes Small Battery Drive II
Страница 2: ......