36 DePuy Synthes User Manual Small Battery Drive II
1.
3.
Spray and wipe.
Spray the device using a
recommended dose of the detergent or neutral pH
enzymatic cleaner (e.g. Steris Prolystica 2X
Concentrate Enzymatic Cleaner 1/8 oz. per gallon).
Leave the solution to interact with the surface of the
device for a minimum of 2 minutes and then wipe it
off (Fig. 2). Follow the enzymatic detergent
manufacturer’s directions for correct temperature,
water quality (i.e. pH, hardness) and
concentration / dilution.
4. Rinse with tap water.
Rinse device with cold tap
water for a minimum of 2 minutes. Use a syringe or
pipette to flush lumens and channels.
5.
Clean with detergent.
Use recommended dose of
the detergent or neutral pH enzymatic cleaner (e.g.
Steris Prolystica 2X Concentrate Enzymatic Cleaner
1/8 oz. per gallon) and clean the device manually
under running warm water for a minimum of 5
minutes. Manipulate all moving parts under running
water. Use a soft-bristled brush and / or soft lint-free
cloth to remove all visible soil and debris.
Follow the enzymatic cleaner or detergent
manufacturer’s instructions for use for correct
temperature, water quality and
concentration / dilution.
6.
Rinse with tap water.
Rinse the device thoroughly
using cool to lukewarm running water for a
minimum of 2 minutes. Use a syringe, pipette or
water jet to flush lumens and channels. Actuate
joints, handles and other movable device features in
order to rinse thoroughly under running water.
7.
Visually inspect device.
Inspect the cannulations,
coupling sleeves, etc. for visible soil. Repeat steps
1–7 until no visible soil remains.
Care and Maintenance
Cleaning and Disinfection
Automated Cleaning Instructions with Manual Pre-Cleaning