Small Battery Drive II User Manual DePuy Synthes 13
Use
Warnings:
•
Do not dismantle, open, short-circuit or shred
batteries.
•
In case of cell leakage, do not allow the leaking
fluid to come into contact with skin or eyes. In
case of contact, wash the affected area with
copious amounts of water and seek medical
advice.
•
Only use the battery for its intended use. Do not use
any battery which is not designed for the equipment.
•
Only insert the battery pack (battery and battery
casing) in the handpiece directly before using the Small
Battery Drive II system. This saves battery energy and
prevents having the need to change it during surgery.
•
Do not apply force to the battery and do not let it fall.
This will destroy it with possible secondary damage.
•
Do not use a faulty or damaged battery, as this may
damage the power tool. Test the status of the battery
using the Universal Battery Charger II.
•
If the drive unit is defective (e.g. short-circuited) do not
insert a battery, as this will blow the internal fuse and
cause damage to the battery. Send the drive unit and
battery to the Depuy Synthes Service Center.
Care and Maintenance
Warnings:
•
Never immerse the handpiece, batteries, battery
casing or attachments in aqueous solutions or in
an ultrasonic bath.
•
Do not use pressurized water as this will cause
damage to the system.
•
The battery must never be autoclaved, washed,
rinsed or dropped. This would destroy the battery
with possible secondary damage.
•
The aseptic transfer is detailed in chapter “Use”.
Alternatively for the Li-Ion battery 532.103 follow
the guidelines provided in the STERRAD/V-PRO
sterilization guide. No other sterilization methods
are allowed.
•
Never expose batteries to temperature over
60 °C/140 °F. The maximum exposure time at
60 °C/140 °F is 72 hours.
Charging, storing and using of Small Battery Drive
II batteries (532.103)
Charging
Warning:
Only use the Depuy Synthes Universal
Battery Charger II (05.001.204) to charge the battery.
Using a charger that does not originate from Depuy
Synthes can damage the battery. The charger should
have a software version of 11.0 or higher. A label
placed on the bottom of the charger enables to
identify the latest software version of the charger.
Precautions:
•
The tool must only be operated with a fully
charged battery. Therefore, ensure that the battery
is charged in good time.
•
Install the battery pack just before use in order to
prevent unwanted discharge of battery capacity.
Furthermore, it is recommended that the battery is
placed into the charger immediately after surgery.
Storage
Warning:
Do not expose batteries to heat or fire.
Avoid storage in direct sunlight.
Precaution:
Do not leave the battery empty of charge
as this will reduce its lifetime or damage the
battery. Always recharge the battery after each use.
Do not store an empty battery as this will shorten
the life span and will not be covered by warranty.
•
When the battery is not being used, always store it in
the Depuy Synthes Universal Battery Charger II and turn
the charging station on. This will avoid the battery from
discharging, and it will be fully charged and ready to
use. Never store the battery in another Depuy Synthes
charger nor a charger from another manufacturer. This
will damage the battery.
•
Keep the batteries and the Universal Battery Charger II
clean and in a cool and dry place.
•
Never store the battery in the battery casing (532.132)
when it is connected to the Small Battery Drive II
handpiece (532.110) as this will discharge the battery.
•
When storing batteries, ensure that they are separately
packed and do not store them with materials that
conduct electricity in order to prevent short-circuit.
This could damage the battery and generate heat
which can cause burns.