J
L
K
Small Battery Drive II User Manual DePuy Synthes 21
Mini Quick Coupling (532.011)
J-Latch Coupling (532.012)
For tools with a Mini Quick or J-Latch coupling shaft.
Mounting and removing the tools
To connect the tool, pull the coupling sleeve back and
then introduce the tool while rotating slightly.
To disconnect, push the coupling sleeve of the
attachment back and remove the tool.
Quick Coupling for Kirschner Wires (532.022)
Kirschner Wires of any length with a diameter of
0.6–3.2 mm can be used with the Quick Coupling for
Kirschner Wires.
1. Adjust the Kirschner Wire diameter according to the
label on the adjusting sleeve
K.
Slightly press the
adjusting sleeve axially against the handpiece and
rotate the sleeve.
2. Apply a slight amount of pressure to insert the
Kirschner Wire from the front into the cannulation
J.
The wire is held automatically.
3. Adjust the working length by pulling on the wire.
4. To affix the wire, pull the tension lever
L
against the
tool with your little finger and ring finger. Only pull
the tension lever against the tool as much as
necessary. The clamping force can be varied by
pulling and releasing the clamping lever.
5. Insert the wire into the bone. Apply the clamping
force as long as the wire is advanced.
6. To adjust the grip on the wire, reduce the clamping
force and move the tool to the desired length.
Reclamp the wire by pulling on the tension lever.