Before initial use, brand-new tools and accessories must
undergo the entire reprocessing process. Completely
remove protective caps and films. To ensure the
longest possible service life, the manufacturer has
partially charged the battery for transport and
storage purposes.
The battery has to be fully charged before first use. This
initial standard charging operation must be carried out
only by the end user (hospital).
Inserting the unsterile battery in the battery casing
Aspetic transfer
The aseptic transfer is detailed below. Alternatively for
the Li-Ion battery 532.103 follow the guidelines provided
in the STERRAD/V-PRO sterilization guide.
To ensure sterility of the battery casing, the battery is
inserted into the battery casing by two people, one of
whom is wearing sterile garments (scrubbed person):
1. The scrubbed person holds the sterile battery asing.
If the casing is not opened, the same person presses
the central button to unlock (Fig. 1), turns the lid
sideways (90°) as indicated by the arrow (Fig. 2) and
pulls to open (Fig. 3). Leave the locking mechanism
swung outward.
2. The scrubbed person places the sterile cover on
the battery casing (Fig. 4) and checks if it is
seated correctly. The sterile cover ensures that
the unsterile battery does not contact the outside
of the sterile casing.
Small Battery Drive II
Use
Figure 4
Figure 3
Figure 1
Figure 2