1
1
1
1
Small Battery Drive II User Manual DePuy Synthes 41
Attachments
All moving parts of all the attachments. Exception: the
Radiolucent Drive (511.30) does not require lubrication.
Chuck (05.001.252–05.001.254)
Lubricate the jaws and toothed rim.
Open and close the drill chuck several times.
Quick Coupling for Kirschner Wires (532.022)
Lubricate the tension lever and clamping mechanism.
Hold the Quick Coupling up and add one drop of oil into
the attachment hole and on the holder of the lever
(Fig. 5).
Move the tension lever several times.
Mini Quick Coupling (532.011)
J-Latch Coupling (532.012)
AO/ASIF Quick Coupling (05.001.250 / 05.001.251)
Quick Coupling for DHS/DCS Triple Reamers
(532.015)
Quick Coupling for Medullary Reaming
(532.017 / 532.018 / 532.019 / 532.020)
Lubricate the unlocking ring. Move it back and forth
several times.
Oscillating Saw Attachment (532.021)
Lubricate the locking mechanism and the saw blade
coupling. Open and close the locking mechanism
several times.
Oscillating Saw Attachment II (532.023)
Lubricate the unlocking sleeve, the tool holder and
the attachment coupling. Move it back and forth
several times.
Large Oscillating Saw Attachment (532.026)
First lubricate and then move all movable parts:
•
Saw blade coupling (slot between the saw blade
coupling and attachment)
•
Locking sleeve of the attachment coupling
(slots on both sides)
•
Coupling pins
•
Opening of the attachment coupling
Figure 6
Figure 5
Содержание DePuy Synthes Small Battery Drive II
Страница 2: ......