Small Battery Drive II User Manual DePuy Synthes 11
Figure 7
Figure 8
Figure 5
Figure 6
3. The circulating person carefully guides the unsterile
battery through the sterile cover (Fig. 5). As an
orientation guide, the two symbols of the battery
and the sterile cover should face each other (Fig. 6).
The same person presses it completely into the
battery casing to ensure a correct seat (Fig. 7).
This person may not contact the outside of the
battery casing.
4. The circulating person grasps the flanges on the
sterile cover and removes it from the battery casing
(Fig. 8).
5. The scrubbed person closes the casing cover from
the outside without contacting the battery or the
inside of the casing. After having closed the casing
cover, turn the lid sideways (90°) until it clicks.
Warnings:
•
Sterilize the sterile cover after each use to ensure
aseptic conditions when inserting the unsterile
battery into the sterile battery casing.
•
The circulating person must avoid all contact with
the outside of the battery casing in order not to
contaminate it. Should the non-sterile battery or
the circulating person’s hand come into contact
with the outside of the battery casing, it must be
cleaned and resterilized before being used in the
operating room.
•
The scrubbed person must not contact the
non-sterile battery or the inside of the battery
casing in order to avoid contamination. Should
the scrubbed person come into contact with
either the non-sterile battery or the inside of
the battery casing, they must be scrubbed again.
Should the battery casing be contaminated, it
must be cleaned and resterilized before being
used in the operating room.
•
If the non-sterile battery contacts the outside
of the battery casing, the battery casing must
be cleaned and resterilized before being used
in the operating room.
•
Do not insert the non-sterile battery into the
battery casing whilst a handpiece is attached.