24 DePuy Synthes User Manual Small Battery Drive II
Attachments
Other Rotating Attachments
Warnings:
•
Grip the coupled Radiolucent Drive tightly when
switching on the power tool, particularly if the
power tool is held face down.
•
Only special 3-flute spiral drill bits can be used.
Your Depuy Synthes representative will provide
you with additional information on which drill
bits can be used.
Precautions:
•
To protect the gears, the Radiolucent Drive is
equipped with a slip clutch that disengages in case
of an overload and emits an audible rattling noise.
•
The following procedures can cause an overload:
–
Correcting the drilling angle when the cutting
edges of the drill bit are completely in the bone.
–
Hitting the nail with the drill bit.
•
Drilling can be continued after having made the
following corrections:
–
Correcting the drilling angle: Remove the drill
bit until the flutes are visible and then restart
the drilling.
–
Hitting a nail: Remove the drill bit until the
flutes are visible and re-aim the drill bit or
exchange the drill bit if necessary.
•
Handle the Radiolucent Drive with great care.
Do not allow contact between the drill bit and the
medullary nail.
•
Depending on the setting of the image intensifier,
a zone may appear in the rear of the Radiolucent
Drive that is not radiolucent. However, this does
not inhibit aiming and working with the device.
Содержание DePuy Synthes Small Battery Drive II
Страница 2: ......