Small Battery Drive II User Manual DePuy Synthes 23
Using the Radiolucent Drive
Before positioning the Radiolucent Drive, align the image
intensifier until the distal locking hole of the medullary
nail is round and easily visible.
After the incision, position the Radiolucent Drive and
center the drill bit tip over the locking hole. On the
monitor of the image intensifier, you can see both the
drill bit and the target rings of the drive.
Swing the drive up and center it precisely so that the drill
bit appears as a round point and the locking hole is
visible around it. The target ring also assists the
centering. The locking hole can now be drilled directly.