STERRAD
®
200 User’s Guide
81
Appendix A.
Specifications
Space Requirements
Sterilizer Size
Width: 915 mm (36 inches)
Depth: 1130 mm (44 ½ inches)
Height: 1754 mm (69 inches)
Weight
475 kg/1047 lbs.
Operation
Electrical Operation
208 VAC, 60 Hz 3 phase WYE, 20 amps NEMA L21-20
receptacle, CBA phase rotation
Ambient Temperature
+18
o
C to +40
o
C (+64
o
F to +103
o
F)
Relative Humidity
10% to 85% up to +30
o
C ambient and linearly decreasing
from 85% at +30
o
C to +70% at +40
o
C noncondensing
Air Exchanges
Minimum 10/hour
Heat Generation
1919 BTU/Cycle
Pollution Degree
2
Installation Category
II
Atmospheric Pressure
700 hPa to 1060 hPa
Altitude
Up to 2000 meters
Содержание ASP STERRAD 200
Страница 1: ...STERRAD 200 Sterilization System User s Guide REF 99712 99712_01 May 2008...
Страница 16: ......
Страница 28: ......
Страница 37: ...Day to Day Operation 4 STERRAD 200 User s Guide 35 STERRAD CycleSure Biological Indicator...
Страница 54: ...4 Day to Day Operation 52 STERRAD 200 User s Guide Figure 19 Cycle In Progress Numerical Display...
Страница 63: ...Day to Day Operation 4 STERRAD 200 User s Guide 61 Cycle Completion Flow Chart...
Страница 74: ......
Страница 82: ......