3
Preparing Items To Be Sterilized
20
STERRAD
®
200 User’s Guide
Items Not Recommended
♦
Items made with copper or copper alloys (such as Monel), should not be
used. Please contact the ASP Customer Care Center at 1-888-STERRAD for
more information.
♦
Instrument mats other than STERRAD Instrument Mats.
♦
Instrument trays other than STERRAD Instrument Trays or APTIMAX
®
Instrument Trays.
♦
Any item that is not completely dry.
♦
Items or materials that absorb liquids.
♦
Items made of materials that contain cellulose, such as cotton, paper or
cardboard, linens, huck towels, gauze sponges, or any item containing wood
pulp.
♦
Paper instrument count sheets or lot stickers.
♦
Liquids and powders.
♦
Items with mated, Nylon
®
surfaces.
♦
Single use items for which the manufacturer does not recommend
resterilization.
♦
Implants for which the manufacturer has not specifically recommended
sterilization in the STERRAD 200 Sterilizer.
♦
Instruments and devices that cannot withstand a vacuum and are labeled for
gravity steam sterilization methods only.
♦
Items whose design permits the surfaces to collapse onto each other unless
some method is used to keep the surfaces separated.
♦
Devices with internal parts, such as sealed bearings, that cannot be immersed
may present difficulties in cleaning and should not be processed in the
STERRAD 200 Sterilizer.
Содержание ASP STERRAD 200
Страница 1: ...STERRAD 200 Sterilization System User s Guide REF 99712 99712_01 May 2008...
Страница 16: ......
Страница 28: ......
Страница 37: ...Day to Day Operation 4 STERRAD 200 User s Guide 35 STERRAD CycleSure Biological Indicator...
Страница 54: ...4 Day to Day Operation 52 STERRAD 200 User s Guide Figure 19 Cycle In Progress Numerical Display...
Страница 63: ...Day to Day Operation 4 STERRAD 200 User s Guide 61 Cycle Completion Flow Chart...
Страница 74: ......
Страница 82: ......