Day-to-Day Operation
4
STERRAD
®
200 User’s Guide
59
4. Push the carriage gently into the sterilizer so that the foot mechanism on the
bottom of the sterilizer contacts and locks into the flange at the bottom of the
carriage. The top front of the carriage rests just inside the chamber and
releases the latch holding the two-tier shelf in the sterilizer.
5. Gently pull the two-tier shelf onto the carriage. Make sure the alignment
wheels are correctly positioned on the rails for smooth movement.
Figure 24. Make sure the carriage is correctly positioned and locked
to the foot mechanism. Pull the two-tier shelf gently onto the carriage.
6. Continue pulling until the two-tier shelf is completely seated on the carriage
and is locked in place.
7. Step on the carriage release and move the two-tier shelf and carriage to the
desired location.
8. Touch Close Door.