Day-to-Day Operation
4
STERRAD
®
200 User’s Guide
43
Figure 12. If skin contact occurs,
immediately flush the area with water.
WARNING! DO NOT REMOVE THE PLASTIC WRAPPER FROM
THE CASSETTE PACKAGE IF THE INDICATOR
STRIP IS RED. RED INDICATES DAMAGE. CALL
THE ASP CUSTOMER CARE CENTER FOR CREDIT.
To insert a cassette, do the following:
1. Confirm that the sterilizer display indicates that a new cassette is needed.
2. Confirm that the chemical indicator strip on the cassette sleeve is NOT red;
red indicates a damaged cassette and concentrated hydrogen peroxide may be
present.
3. Confirm that the cassette expiration date has not passed.
Note:
The system considers the cassette expired 8 days after
insertion regardless of the printed expiration date. The
cassette is ejected at that time. Once the cassette has been
removed, even if it was not used, the barcode is rendered
unreadable and the cassette cannot be reinserted.