Troubleshooting
6
STERRAD
®
200 User’s Guide
79
Displayed Messages
Printed Messages
Condition/Action
Call ASP Technical Service
needed to run a cycle. Call the ASP
Customer Care Center.
Temperature Out Of Range
Temperature Out Of
Range (System
temperature parameters
are included on printout.)
The door, chamber, vaporizer, or tube
temperatures are above or below the
operating temperatures. Call the ASP
Customer Care Center.
Timeout Failure
Timeout Failure
A timeout event occurred when the
sterilizer was unable to complete a task
within a preset timeout period. Run the
cycle again, if the timeout problem
persists, call the ASP Customer Care
Center.
Vacuum System Failure
Vacuum System Failure
(Vacuum system
parameters are included
on printout.)
The load may contain too much
moisture contain cellulose material or
there may be too many packages in the
chamber. If problem persists, start a
cycle with an empty chamber. If the
vacuum stage is still not successful,
call the ASP Customer Care Center.
Warming Up
Please Wait
No Printout
You must wait until the system warms
up to run a cycle.
Содержание ASP STERRAD 200
Страница 1: ...STERRAD 200 Sterilization System User s Guide REF 99712 99712_01 May 2008...
Страница 16: ......
Страница 28: ......
Страница 37: ...Day to Day Operation 4 STERRAD 200 User s Guide 35 STERRAD CycleSure Biological Indicator...
Страница 54: ...4 Day to Day Operation 52 STERRAD 200 User s Guide Figure 19 Cycle In Progress Numerical Display...
Страница 63: ...Day to Day Operation 4 STERRAD 200 User s Guide 61 Cycle Completion Flow Chart...
Страница 74: ......
Страница 82: ......