4
Day-to-Day Operation
44
STERRAD
®
200 User’s Guide
4. Remove the plastic wrapping from the cassette sleeve. Do NOT remove the
cassette from the remaining cardboard sleeve.
5. Orient the arrow so that the top of the cassette sleeve is pointing away from
you.
6. Hold the cardboard sleeve by its edges and insert it into the sterilizer.
7. Firmly push the cardboard sleeve in until it can go no further. If positioned
properly, the cassette moves into place.
Figure 13. Insert the cassette and sleeve into the sterilizer.
This illustration shows the printer on the input side. Your system
may be different.
8. If the cardboard sleeve is not pushed in all the way, the monitor continues to
read INSERT CASSETTE.
9. If the cardboard sleeve is properly positioned, the cassette is automatically
accepted and positioned for use by the sterilizer. The display reads
CASSETTE ACCEPTED.
10. If the cassette is not accepted because the barcode cannot be read or the
cassette has expired, PLEASE REMOVE CASSETTE is displayed. Should
this occur, remove that cassette and insert a valid one.
Note:
The system microprocessor monitors the status of the
cassette and informs the operator when the cassette is
empty or expired. Empty or expired cassettes must be
replaced prior to starting a cycle.
Содержание ASP STERRAD 200
Страница 1: ...STERRAD 200 Sterilization System User s Guide REF 99712 99712_01 May 2008...
Страница 16: ......
Страница 28: ......
Страница 37: ...Day to Day Operation 4 STERRAD 200 User s Guide 35 STERRAD CycleSure Biological Indicator...
Страница 54: ...4 Day to Day Operation 52 STERRAD 200 User s Guide Figure 19 Cycle In Progress Numerical Display...
Страница 63: ...Day to Day Operation 4 STERRAD 200 User s Guide 61 Cycle Completion Flow Chart...
Страница 74: ......
Страница 82: ......