STERRAD
®
200 User’s Guide
17
Chapter 3.
Preparing Items To Be
Sterilized
Overview
The STERRAD
®
200 Sterilizer can process many of the items you commonly
sterilize as well as instruments that are sensitive to heat and moisture. However,
there are a few important exceptions. Please review the “How to Determine What
Can be Sterilized in the STERRAD 200 System” foldout page contained in this
chapter. It contains details on recommended materials and lumen sizes.
Indications for Use
The STERRAD 200 Sterilizer is designed for sterilization of both metal and
nonmetal medical devices at low temperatures. Because the cycle operates within
a dry environment and at low temperatures, it is especially suitable for
instruments sensitive to heat and moisture. The sections that follow contain more
information on the types of items and materials that can be sterilized in the
STERRAD 200 Sterilizer.
•
The STERRAD 200 Sterilizer can sterilize instruments having diffusion-
restricted spaces, such as the hinged portion of forceps and scissors.
•
Teflon
®
or polyethylene lumened instruments with inside diameters of 6 mm
or larger and lengths of 310 mm or shorter can be processed in the
STERRAD 200 Sterilizer.
•
Medical devices with only a single stainless steel lumen with:
Содержание ASP STERRAD 200
Страница 1: ...STERRAD 200 Sterilization System User s Guide REF 99712 99712_01 May 2008...
Страница 16: ......
Страница 28: ......
Страница 37: ...Day to Day Operation 4 STERRAD 200 User s Guide 35 STERRAD CycleSure Biological Indicator...
Страница 54: ...4 Day to Day Operation 52 STERRAD 200 User s Guide Figure 19 Cycle In Progress Numerical Display...
Страница 63: ...Day to Day Operation 4 STERRAD 200 User s Guide 61 Cycle Completion Flow Chart...
Страница 74: ......
Страница 82: ......