Day-to-Day Operation
4
STERRAD
®
200 User’s Guide
51
Watching a Cycle
You can monitor the progress of a cycle by watching the display; it indicates the
phase of the cycle and certain process parameters. A long beep signals that the
cycle is complete.
The display indicates the status of the unit at all times: the current stage of the
sterilization cycle; the temperature in the chamber; the pressure in the chamber;
the cycle start time, and the cycle estimated end time. You can change the display
to show a graph or a numerical readout of the cycle. Each load goes through nine
consecutive stages: vacuum, pre-plasma, injection, diffusion, plasma, injection,
diffusion, plasma, and vent.
Figure 18. Cycle In Progress–Graph Display. Your display may look slightly different.
Содержание ASP STERRAD 200
Страница 1: ...STERRAD 200 Sterilization System User s Guide REF 99712 99712_01 May 2008...
Страница 16: ......
Страница 28: ......
Страница 37: ...Day to Day Operation 4 STERRAD 200 User s Guide 35 STERRAD CycleSure Biological Indicator...
Страница 54: ...4 Day to Day Operation 52 STERRAD 200 User s Guide Figure 19 Cycle In Progress Numerical Display...
Страница 63: ...Day to Day Operation 4 STERRAD 200 User s Guide 61 Cycle Completion Flow Chart...
Страница 74: ......
Страница 82: ......