STERRAD
®
200 User’s Guide
77
Chapter 6.
Troubleshooting
Overview
The STERRAD
®
200 Sterilizer is a relatively trouble-free device, requiring only
routine maintenance and care in load preparation to help prevent system
cancellations.
WARNING! YOU MUST WEAR CHEMICAL RESISTANT LATEX,
PVC (VINYL), OR NITRILE GLOVES WHEN
REMOVING A LOAD AFTER A CYCLE
CANCELLATION.
Rebooting the System
Rebooting is used in certain troubleshooting procedures. When power is turned
off then back on, it causes the computer in the sterilizer to reload the software
control program automatically; this action “reboots” the sterilizer. (Rebooting is
used in certain troubleshooting procedures.) The display indicates sterilizer status
after power has been reapplied.
The Power ON-OFF switch is located on the service side of the sterilizer (this is
to the right as you face the input side of the sterilizer). If instructions indicate that
you should reboot the system, flip the switch to the OFF position. Wait a few
seconds for the system to completely power down, then flip the switch to the ON
position. The display indicates sterilizer status after power has been reapplied.
Содержание ASP STERRAD 200
Страница 1: ...STERRAD 200 Sterilization System User s Guide REF 99712 99712_01 May 2008...
Страница 16: ......
Страница 28: ......
Страница 37: ...Day to Day Operation 4 STERRAD 200 User s Guide 35 STERRAD CycleSure Biological Indicator...
Страница 54: ...4 Day to Day Operation 52 STERRAD 200 User s Guide Figure 19 Cycle In Progress Numerical Display...
Страница 63: ...Day to Day Operation 4 STERRAD 200 User s Guide 61 Cycle Completion Flow Chart...
Страница 74: ......
Страница 82: ......