6
Troubleshooting
78
STERRAD
®
200 User’s Guide
Using the Message Table
To use this table, look at the message displayed and then find the same message
in the table (it is in alphabetical order). The table shows the message, the printout
associated with the displayed message, and suggests some actions to take. If you
touch
Help
when any message is displayed, you receive additional solutions and
actions to take. You may receive similar messages during different stages of the
system cycle. The action you are to take is usually the same regardless of the
cycle stage.
If your question was not answered either by the table below or by using the
“Help” display, call the ASP Customer Care Center for more information.
Message Table
Displayed Messages
Printed Messages
Condition/Action
Delivery Subsystem Failure
Delivery Subsystem
Failure
The system was unable to successfully
complete the injection. Call the ASP
Customer Care Center.
H
2
O
2
Delivery Failure
H
2
O
2
Delivery Failure
(System parameters are
included on printout.)
Acceptable hydrogen peroxide
concentration was not detected. Insert
a new cassette and run the cycle again.
If the problem persists, call the ASP
Customer Care Center.
Insert New Cassette
No Printout
Insert a new cassette and restart the
sterilizer.
Operator Cancellation
Operator Cancellation
The operator has cancelled the cycle.
Restart the sterilizer.
Power Failure
Power Failure
The power was turned off during a
cycle. Restart the sterilizer
RF Subsystem Failure
RF Subsystem Failure
(RF system parameters
are included on printout.)
WARNING!
If plasma failed to light,
please use extra caution when handling
load.
Make sure the shelf or other metal
objects are not in contact with the
electrode, door, or chamber walls.
Repackage and reprocess load. If
problem persists, call the ASP
Customer Care Center.
Temperature Has Not Risen
No Printout
The sterilizer is not at the temperature
Содержание ASP STERRAD 200
Страница 1: ...STERRAD 200 Sterilization System User s Guide REF 99712 99712_01 May 2008...
Страница 16: ......
Страница 28: ......
Страница 37: ...Day to Day Operation 4 STERRAD 200 User s Guide 35 STERRAD CycleSure Biological Indicator...
Страница 54: ...4 Day to Day Operation 52 STERRAD 200 User s Guide Figure 19 Cycle In Progress Numerical Display...
Страница 63: ...Day to Day Operation 4 STERRAD 200 User s Guide 61 Cycle Completion Flow Chart...
Страница 74: ......
Страница 82: ......